Micro Finite Element models of the vertebral body: Validation of local displacement predictions.

Maria Cristiana Costa, Gianluca Tozzi, Luca Cristofolini, Valentina Danesi, Marco Viceconti, Enrico Dall'Ara

Open access · goldAbstract read
In one paragraph

Article in PloS one, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 23 papers.

0numbers the graph read from it
0cells of the map it votes in
23citing papers in PubMed
7.7field-weighted citation impact, top 2% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

23 citing papers in PubMed, 69 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Review
  9. Article
  10. Review
  11. Properties of PMMA end cap holders affect FE stiffness predictions of vertebral specimens.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2021
    Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Efficient materially nonlinearBiomechanics and modeling in mechanobiology · 2020
    Article
  18. Article
  19. Article
  20. Bone Damage Evolution Around Integrated Metal Screws Using X-Ray Tomography -Frontiers in bioengineering and biotechnology · 2020
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 3 institutions in 2 countries.

Maria Cristiana CostaDepartment of Oncology and Metabolism, University of Sheffield, Sheffield, United Kingdom.
Gianluca TozziZeiss Global Centre, School of Engineering, University of Portsmouth, Portsmouth, United Kingdom.
Luca CristofoliniSchool of Engineering and Architecture, Alma Mater Studiorum-Università di Bologna, Bologna, Italy.
Valentina DanesiSchool of Engineering and Architecture, Alma Mater Studiorum-Università di Bologna, Bologna, Italy.
Marco VicecontiINSIGNEO Institute for In Silico Medicine, University of Sheffield, Sheffield, United Kingdom.
Enrico Dall'AraDepartment of Oncology and Metabolism, University of Sheffield, Sheffield, United Kingdom.ORCID http://orcid.org/0000-0003-1471-5077
Insigneo · GBUniversity of Bologna · ITUniversity of Portsmouth · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The estimation of local and structural mechanical properties of bones with micro Finite Element (microFE) models based on Micro Computed Tomography images depends on the quality bone geometry is captured, reconstructed and modelled. The aim of this study was to validate microFE models predictions of local displacements for vertebral bodies and to evaluate the effect of the elastic tissue modulus on model's predictions of axial forces. Four porcine thoracic vertebrae were axially compressed in situ, in a step-wise fashion and scanned at approximately 39μm resolution in preloaded and loaded conditions. A global digital volume correlation (DVC) approach was used to compute the full-field displacements. Homogeneous, isotropic and linear elastic microFE models were generated with boundary conditions assigned from the interpolated displacement field measured from the DVC. Measured and predicted local displacements were compared for the cortical and trabecular compartments in the middle of the specimens. Models were run with two different tissue moduli defined from microindentation data (12.0GPa) and a back-calculation procedure (4.6GPa). The predicted sum of axial reaction forces was compared to the experimental values for each specimen. MicroFE models predicted more than 87% of the variation in the displacement measurements (R2 = 0.87-0.99). However, model predictions of axial forces were largely overestimated (80-369%) for a tissue modulus of 12.0GPa, whereas differences in the range 10-80% were found for a back-calculated tissue modulus. The specimen with the lowest density showed a large number of elements strained beyond yield and the highest predictive errors. This study shows that the simplest microFE models can accurately predict quantitatively the local displacements and qualitatively the strain distribution within the vertebral body, independently from the considered bone types.

Indexed as

Finite Element AnalysisAnimalsElastic ModulusLumbar VertebraeStress, MechanicalSwineThoracic Vertebrae

Identifiers

PMID28700618
PMCPMC5507408
OpenAlexW2650556801

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.