Evidence map›Paper›PMID 28622370›Full record

ArticlePloS one2017

Evaluation of the performance of Abbott m2000 and Roche COBAS Ampliprep/COBAS Taqman assays for HIV-1 viral load determination using dried blood spots and dried plasma spots in Kenya.

Clement Zeh, Kenneth Ndiege, Seth Inzaule, Rebecca Achieng, John Williamson, Joy Chih-Wei Chang, Dennis Ellenberger, John Nkengasong

Open access · goldAbstract readEvaluation Study
In one paragraph

Article in PloS one, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 25 papers, 3 of them syntheses that pooled it.

0numbers the graph read from it
0cells of the map it votes in
25citing papers in PubMed, 3 pooled it
3.1field-weighted citation impact, top 8% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

25 citing papers in PubMed, 3 syntheses or guidelines pooled it, 38 citations in OpenAlex.

  1. Pooled it
  2. Pooled it
  3. Pooled it
  4. Review
  5. Article
  6. Article
  7. Review
  8. Review
  9. Article
  10. Review
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Cost and Impact of Dried Blood Spot Versus Plasma Separation Card for Scale-up of Viral Load Testing in Resource-limited Settings.Clinical infectious diseases : an official publication of the Infectious Diseases Society of America · 2020
    Article
  18. Article
  19. Article
  20. Observational
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 2 countries.

Clement ZehDivision of HIV/AIDS Prevention, U.S. Centers for Disease Control and Prevention, Kisumu, Kenya.ORCID http://orcid.org/0000-0002-5023-0336
Kenneth NdiegeCentre for Global Health Research, Kenya Medical Research Institute, Kisumu, Kenya.
Seth InzauleCentre for Global Health Research, Kenya Medical Research Institute, Kisumu, Kenya.
Rebecca AchiengCentre for Global Health Research, Kenya Medical Research Institute, Kisumu, Kenya.
John WilliamsonDivision of HIV/AIDS Prevention, U.S. Centers for Disease Control and Prevention, Kisumu, Kenya.
Joy Chih-Wei ChangInternational Laboratory Branch, Division of Global HIV/AIDS, Center for Global Health, U.S. Centers for Disease Control and Prevention, Atlanta, Georgia, United States of America.
Dennis EllenbergerInternational Laboratory Branch, Division of Global HIV/AIDS, Center for Global Health, U.S. Centers for Disease Control and Prevention, Atlanta, Georgia, United States of America.
John NkengasongInternational Laboratory Branch, Division of Global HIV/AIDS, Center for Global Health, U.S. Centers for Disease Control and Prevention, Atlanta, Georgia, United States of America.
Centers for Disease Control and Prevention · USKenya Medical Research Institute · KECenters for Disease Control and Prevention · KE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundRoutine HIV viral load testing is not widely accessible in most resource-limited settings, including Kenya. To increase access to viral load testing, alternative sample types like dried blood spots (DBS), which overcome the logistic barriers associated with plasma separation and cold chain shipment need to be considered and evaluated. The current study evaluated matched dried blood spots (DBS) and dried plasma spots (DPS) against plasma using the Abbott M 2000 (Abbott) and Roche Cobas Ampliprep/Cobas TaqMan (CAP/CTM) quantitative viral load assays in western Kenya.

methodsMatched plasma DBS and DPS were obtained from 200 HIV-1 infected antiretroviral treatment (ART)-experienced patients attending patient support centers in Western Kenya. Standard quantitative assay performance parameters with accompanying 95% confidence intervals (CI) were assessed at the assays lower detection limit (400cps/ml for CAP/CTM and 550cps/ml for Abbott) using SAS version 9.2. Receiver operating curves (ROC) were further used to assess viral-load thresholds with best assay performance (reference assay CAP/CTM plasma).

resultsUsing the Abbott test, the sensitivity and specificity, respectively, for DPS were (97.3%, [95%CI: 93.2-99.2] and 98.1% [95%CI: 89.7-100]) and those for DBS (93.9% [95%CI: 88.8-97.2] and 88.0% [95%CI: 82.2-92.4]). The correlation and agreement using paired plasma and DPS/DBS were strong, with r2 = 90.5 and rc = 68.1. The Bland-Altman relative percent change was 95.3 for DPS, (95%CI: 90.4-97.7) and 73.6 (95%CI: 51.6-86.5) for DBS. Using the CAP/CTM assay, the sensitivity for DBS was significantly higher compared to DPS (100.0% [95% CI: 97.6-100.0] vs. 94.7% [95%CI: 89.8-97.7]), while the specificity for DBS was lower: 4%, [95% CI: 0.4-13.7] compared to DPS: 94.0%, [95% CI: 83.5-98.7]. When compared under different clinical relevant thresholds, the accuracy for the Abbott assay was 95% at the 1000cps/ml cut-off with a sensitivity and specificity of 96.6% [95% CI 91.8-98.7] and 90.4% [95% CI 78.2-96.4] respectively. The optimum threshold was at 3000 cps/ml with an accuracy of 95.5%, sensitivity and specificity of 94.6% [95%CI 89.3-97.5] and 98.1% [95%CI 88.4-99.9]) respectively. The best threshold for CAP/CTM was at 4000 copies /mL, with 92.5% accuracy (sensitivity of 96.0% [95%CI 91.0-98.3] and specificity of 82.7% [95%CI 69.2-91.3]).

conclusionsThere was similar performance between matched DBS, DPS and plasma using the Abbott test, and good correlation for matched DPS and plasma using the CAPCTM test. The findings suggest that DBS and DPS may be reliably used as alternative specimens to plasma to measure HIV-1 VL using Abbott, and DPS may be reliably used with CAP/CTM in resource-limited settings.

Indexed as

Dried Blood Spot TestingHIV-1Polymerase Chain ReactionViral LoadFemaleHIV InfectionsHumansKenyaMaleSensitivity and Specificity

Identifiers

PMID28622370
PMCPMC5473550
OpenAlexW2626621259

What OpenQuestion holds

Textmetadata
LicenceCC0
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.