Trial reportJournal of diabetes science and technology2017
Accuracy and User Performance Evaluation of a New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device.
Trial report in Journal of diabetes science and technology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 13 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
13 citing papers in PubMed.
- Evaluation of the Bluetooth-enabled Scanbo device for point-of-care measurement of blood pressure and blood glucose: A cross-sectional pilot study in an urban slum of Bengaluru.Future healthcare journal · 2025Article
- Telemonitoring With a Connected Glucose Meter Improves Glycemia Among People With Insulin-Treated Type 2 Diabetes.Journal of diabetes science and technology · 2023Article
- The effect of type-2 diabetes conditions on neutrophil rolling adhesion.BMC research notes · 2022Article
- Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.Journal of diabetes science and technology · 2022Article
- Smartphone-based mobile biosensors for the point-of-care testing of human metabolites.Materials today. Bio · 2022Review
- Accuracy and User Performance of a New Blood Glucose Monitoring System.Journal of diabetes science and technology · 2021Article
- New and Emerging Technologies in Type 1 Diabetes.Endocrinology and metabolism clinics of North America · 2020Review
- Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.Journal of diabetes science and technology · 2019Article
- Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.Journal of diabetes science and technology · 2019Article
- User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2018Article
- Improved Glycemic Control in a Patient Group Performing 7-Point Profile Self-Monitoring of Blood Glucose and Intensive Data Documentation: An Open-Label, Multicenter, Observational Study.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2017Article
- Analysis of "Accuracy and User Performance Evaluation of a New, Wireless-Enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device".Journal of diabetes science and technology · 2017Article
- Mobile technology for self-monitoring of blood glucose among patients with type 2 diabetes mellitus.mHealth · 2017Article
Corrections and comments
- Commented on by
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe new Contour
methodsThe accuracy of the new BGMS was assessed in 2 studies according to ISO 15197:2013 criteria. In Study 1 (laboratory study), fingertip capillary blood samples from 100 subjects were tested in duplicate using 3 test strip lots. In Study 2 (clinical study), 134 subjects with type 1 or type 2 diabetes enrolled at 2 clinical sites. BGMS results and YSI analyzer (YSI) reference results were compared for fingertip blood obtained by untrained subjects' self-testing and for study staff-obtained fingertip, subject palm, and venous results.
resultsIn Study 1, 99.0% (594/600) of combined results for all 3 test strip lots fulfilled ISO 15197:2013 Section 6.3 accuracy criteria. In Study 2, 99.2% (133/134) of subject-obtained capillary fingertip results, 99.2% (133/134) of study staff-obtained fingertip results, 99.2% (125/126) of subject-obtained palm results, and 100% (132/132) of study staff-obtained venous results met ISO 15197:2013 Section 8 accuracy criteria. Moreover, 95.5% (128/134) of subject-obtained fingertip self-test results were within ±10 mg/dl (±0.6 mmol/L) or ±10% of the YSI reference result. Questionnaire results showed that most subjects found the BGMS easy to use.
conclusionsThe BGMS exceeded ISO 15197:2013 accuracy criteria both in the laboratory and in a clinical setting when used by untrained subjects with diabetes.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials