ArticleThe British journal of radiology2017
Quantification of radiation dose reduction by reducing z-axis coverage in 320-detector coronary CT angiography.
Article in The British journal of radiology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Impact of an ultra-low dose unenhanced planning scan on CT coronary angiography scan length and effective radiation dose.BJR open · 2022Article
- Potential increase in radiation-induced DNA double-strand breaks with higher doses of iodine contrast during coronary CT angiography.Medical physics · 2021Article
- Reduced Axial Scan Length Coronary Calcium Scoring Reduces Radiation Dose and Provides Adequate Clinical Decision-Making Before Coronary CT Angiography.Tomography (Ann Arbor, Mich.) · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo quantify the radiation dose reduction achievable by minimizing z-axis coverage in 320-detector coronary CT angiography (CCTA).
methodsWe retrospectively reviewed 130 CCTAs performed on 320-detector CT that offers up to 16 cm z-axis coverage (adjustable in 2-cm increments), allowing complete coverage of the heart in a single gantry rotation. For each CT, we obtained the radiation dose [CT dose index and dose-length product (DLP)], measured the z-axis field of view and measured the craniocaudal cardiac size (distance from the left main coronary artery to the cardiac apex). We calculated the radiation dose savings achievable by reducing the z-axis coverage to the minimum necessary to cover the heart using 320 × 0.5-mm (maximum 16 cm) and 256 × 0.5-mm (maximum 12.8 cm) detector collimations.
resultsResults are expressed as mean ± standard deviation. The mean craniocaudal cardiac size was 10.5 ± 1.0 cm, with 85% (n = 112) of CCTAs performed with 16 cm of z-axis coverage. The mean DLP was 417.6 ± 182.4 mGy cm, with the mean DLP saving achievable using the minimum z-axis coverage required to completely image the heart being 96.2 ± 47.4 mGy cm, an average dose reduction of 26.9 ± 7.0%. z-axis coverage of ≤12 cm was adequate for 92% and 12.8 cm for 98% of subjects.
conclusionUsing the minimal z-axis coverage to adequately image the heart is a simple step that can reduce the DLP in 320-detector CCTA by approximately 27%. z-axis coverage of ≤12 cm is adequate for 92%, 12.8 cm for 98% and 14 cm for 100% of patients undergoing CCTA. Advances in knowledge: Reducing z-axis coverage in 320-detector CCTA can reduce DLP by approximately 27%.