Evidence map›Paper›PMID 28604183›Full record

ArticleInterventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences2017

Treatment of an internal carotid artery aneurysm with a flow diverter through a double lumen balloon catheter.

José M Amorim, Santiago Rosati, Ronit Agid, Vítor Mendes Pereira, Timo Krings

Abstract readCase Reports
In one paragraph

Article in Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Deployment of low-profile flow-diverter through a double lumen balloon catheter (remo-diversion technique) to treat recurrent middle cerebral artery bifurcation aneurysm: A technical note.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2024
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

José M Amorim1 Neuroradiology Department, Hospital de Braga, Braga, Portugal.
Santiago Rosati2 Neuroradiology Department, Clinical San Carlos Hospital, Madrid, Spain.
Ronit Agid3 Division of Neuroradiology, Department of Medical Imaging, University of Toronto, Toronto Western Hospital, University Health Network, Canada.
Vítor Mendes Pereira3 Division of Neuroradiology, Department of Medical Imaging, University of Toronto, Toronto Western Hospital, University Health Network, Canada.
Timo Krings3 Division of Neuroradiology, Department of Medical Imaging, University of Toronto, Toronto Western Hospital, University Health Network, Canada.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background Double lumen balloon catheters (DLBCs) are currently used in the treatment of intracranial aneurysms, especially when involving balloon or stent-assisted coiling. The existing DLBCs allow the delivery of self-expandable stents but do not offer the possibility to deploy flow-diverters. Despite the increasing use and success of flow-diverters, there have been numerous reports of procedural complications such as early in-stent thrombosis or delayed distal embolization. It seems that these complications can be avoided by correct stent positioning and adequate wall apposition, achieved either by manoeuvres with the microguidewire and/or microcatheter or by performing balloon angioplasty following an exchange guidewire manoeuvre. Objective Report the use of a new DLBC able to deliver a flow-diverter. Methods A 41-year-old woman presented to our hospital with binocular horizontal diplopia for two weeks and reduced visual acuity. A left internal carotid artery aneurysm involving the cavernous and ophthalmic segments was found, with a maximum height of 19 mm and a broad 8 mm neck. It presented extra- and intra-dural components and the parent vessel was significantly narrowed. A decision was made to perform endovascular treatment of the aneurysm with placement of a flow diverter through a DLBC. Results Patency and adequate expansion of the flow diverter with evident intra-aneurysmal contrast stasis was observed in the final angiogram. No peri-procedural complications were observed. Conclusion This is a technical note demonstrating the feasibility of a new device to deploy a flow diverter, aiming to improve wall apposition and stent configuration without the need of additional devices or exchange manoeuvres.

Indexed as

Carotid Artery, InternalAdultAngiography, Digital SubtractionAngioplasty, BalloonCerebral AngiographyEmbolization, TherapeuticFemaleHumansIntracranial AneurysmStentsDouble lumen balloon catheterflow diverterintracranial aneurysm

Identifiers

PMID28604183
PMCPMC5490856

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.