Comparison of the biometric measurements calculated with Zeiss IOL-Master and WaveLight OB-820.

Georgios Labiris, Panagiota Ntonti, Konstantinos Ntitsos, Andreas Katsanos, Haris Sideroudi, Vasilios Kozobolis

Abstract read
In one paragraph

Article in Clinical ophthalmology (Auckland, N.Z.), 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Observational
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Georgios LabirisDepartment of Ophthalmology, University Hospital of Alexandroupolis.
Panagiota NtontiEye Institute of Thrace, Alexandroupolis, Greece.
Konstantinos NtitsosEye Institute of Thrace, Alexandroupolis, Greece.
Andreas KatsanosEye Institute of Thrace, Alexandroupolis, Greece.
Haris SideroudiEye Institute of Thrace, Alexandroupolis, Greece.
Vasilios KozobolisDepartment of Ophthalmology, University Hospital of Alexandroupolis.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe primary objective of this study was to investigate the level of agreement between IOL-Master and OB-820 ocular biometers. MATERIALS AND

methodsIn this prospective randomized case series, we measured the anterior chamber depth (ACD), the axial length (AL), the corneal radii (R1, R2), the ratio of mean corneal radius and spherical equivalent and the corneal astigmatism (cylinder [Cyl]) before and after cataract extraction surgery.

resultsSignificant differences between pre- and postoperative data were observed for ACD for both biometric devices (

conclusionIn clinical practice, the IOL-master and OB-820 should not be used interchangeably due to discrepancy in the important AL parameter. Both biometers may provide consistent results regarding Cyl, ACD and R1.

Indexed as

biometrycataractIOL-Masterlevel of agreementOB-820