ArticleThe bone & joint journal2017
Risk factors for failure of the 36 mm metal-on-metal Pinnacle total hip arthroplasty system: a retrospective single-centre cohort study.
Article in The bone & joint journal, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed.
- Findings From 333 Retrieved MoM Total Hip Arthroplasty Components of a Single Design Reveal Modular Interface Corrosion as Most Common Feature.Journal of the American Academy of Orthopaedic Surgeons. Global research & reviews · 2026Article
- Case report: Formation and recurrence of inflammatory pseudotumor after metal-on-metal hip arthroplasty.Frontiers in medicine · 2024Article
- Perioperative Complications after Hip and Knee Revision Arthroplasty in the over 80 Years Old Population: A Retrospective Observational Case-Control Study.Journal of clinical medicine · 2023Article
- How much does a Medical and Healthcare Products Regulatory Agency medical device alert for metal-on-metal hip arthroplasty patients really cost?Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2022Article
- Has the threshold for revision surgery for adverse reactions to metal debris changed in metal-on-metal hip arthroplasty patients? A cohort study of 239 patients using an adapted risk-stratification algorithm.Acta orthopaedica · 2019Article
- The current role of robotics in total hip arthroplasty.EFORT open reviews · 2019Review
- No Threshold Exists for Recommending Revision Surgery in Metal-on-Metal Hip Arthroplasty Patients With Adverse Reactions to Metal Debris: A Retrospective Cohort Study of 346 Revisions.The Journal of arthroplasty · 2019Article
- Wear performance of retrieved metal-on-metal Pinnacle hip arthroplasties implanted before and after 2007.Bone & joint research · 2018Article
- What is appropriate surveillance for metal-on-metal hip arthroplasty patients?Acta orthopaedica · 2018Article
- The risk of cardiac failure following metal-on-metal hip arthroplasty.The bone & joint journal · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
Abstract
aimsTo determine ten-year failure rates following 36 mm metal-on-metal (MoM) Pinnacle total hip arthroplasty (THA), and identify predictors of failure. PATIENTS AND
methodsWe retrospectively assessed a single-centre cohort of 569 primary 36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012 according to Medicines and Healthcare Products Regulation Agency recommendations. All-cause failure rates (all-cause revision, and non-revised cross-sectional imaging failures) were calculated, with predictors for failure identified using multivariable Cox regression.
resultsFailure occurred in 97 hips (17.0%). The ten-year cumulative failure rate was 27.1% (95% confidence interval (CI) 21.6 to 33.7). Primary implantation from 2006 onwards (hazard ratio (HR) 4.30; 95% CI 1.82 to 10.1; p = 0.001) and bilateral MoM hip arthroplasty (HR 1.59; 95% CI 1.03 to 2.46; p = 0.037) predicted failure. The effect of implantation year on failure varied over time. From four years onwards following surgery, hips implanted since 2006 had significantly higher failure rates (eight years 28.3%; 95% CI 23.1 to 34.5) compared with hips implanted before 2006 (eight years 6.3%; 95% CI 2.4 to 15.8) (HR 15.2; 95% CI 2.11 to 110.4; p = 0.007).
conclusionWe observed that 36 mm MoM Pinnacle THAs have an unacceptably high ten-year failure rate, especially if implanted from 2006 onwards or in bilateral MoM hip patients. Our findings regarding implantation year and failure support recent concerns about the device manufacturing process. We recommend all patients undergoing implantation since 2006 and those with bilateral MoM hips undergo regular investigation, regardless of symptoms. Cite this article: