ArticleThe British journal of radiology2017
Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit.
Article in The British journal of radiology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Radiochromic film dosimetry in radiotherapy: a survey of current practice in the United Kingdom.The British journal of radiology · 2024Review
- Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom.Journal of applied clinical medical physics · 2022Article
- Evaluation of the RUBY modular QA phantom for planar and non-coplanar VMAT and stereotactic radiations.Journal of applied clinical medical physics · 2020Article
- Development of a dedicated phantom for multi-target single-isocentre stereotactic radiosurgery end to end testing.Journal of applied clinical medical physics · 2018Article
- The role of dosimetry audit in achieving high quality radiotherapy.Physics and imaging in radiation oncology · 2018Article
- Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.Physics and imaging in radiation oncology · 2018Review
- Validation of the Fabricated Cast Nylon Head Phantom for Stereotactic Radiosurgery End-to-End Test using Alanine Dosimeter.Journal of medical physicsArticle
- Historical Progress of Stereotactic Radiation Surgery.Journal of medical physicsReview
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo adapt and validate an anthropomorphic head phantom for use in a cranial radiosurgery audit.
methodsTwo bespoke inserts were produced for the phantom: one for providing the target and organ at risk for delineation and the other for performing dose measurements. The inserts were tested to assess their positional accuracy. A basic treatment plan dose verification with an ionization chamber was performed to establish a baseline accuracy for the phantom and beam model. The phantom and inserts were then used to perform dose verification measurements of a radiosurgery plan. The dose was measured with alanine pellets, EBT extended dose film and a plastic scintillation detector (PSD).
resultsBoth inserts showed reproducible positioning (±0.5 mm) and good positional agreement between them (±0.6 mm). The basic treatment plan measurements showed agreement to the treatment planning system (TPS) within 0.5%. Repeated film measurements showed consistent gamma passing rates with good agreement to the TPS. For 2%-2 mm global gamma, the mean passing rate was 96.7% and the variation in passing rates did not exceed 2.1%. The alanine pellets and PSD showed good agreement with the TPS (-0.1% and 0.3% dose difference in the target) and good agreement with each other (within 1%).
conclusionThe adaptations to the phantom showed acceptable accuracies. The presence of alanine and PSD do not affect film measurements significantly, enabling simultaneous measurements by all three detectors. Advances in knowledge: A novel method for thorough end-to-end test of radiosurgery, with capability to incorporate all steps of the clinical pathway in a time-efficient and reproducible manner, suitable for a national audit.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials