Assessment of female breast dose for thoracic cone-beam CT using MOSFET dosimeters.
Article in Oncotarget, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Evaluation of registration accuracy using short-arc cone-beam computed tomography for image-guided radiotherapy in breast cancer.Journal of applied clinical medical physics · 2026Article
- The performance of a new type accelerator uRT-linac 506c evaluated by a quality assurance automation system.Journal of applied clinical medical physics · 2024Article
- Monte Carlo-based software for 3D personalized dose calculations in image-guided radiotherapy.Physics and imaging in radiation oncology · 2022Article
- Impact of CBCT frequency on target coverage and dose to the organs at risk in adjuvant breast cancer radiotherapy.Scientific reports · 2021Article
- Application of 3D-print silica bolus for nasal NK/T-cell lymphoma radiation therapy.Journal of radiation research · 2020Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo assess the breast dose during a routine thoracic cone-beam CT (CBCT) check with the efforts to explore the possible dose reduction strategy. MATERIALS AND
methodsMetal oxide semiconductor field-effect transistor (MOSFET) dosimeters were used to measure breast surface doses during a thorax kV CBCT scan in an anthropomorphic phantom. Breast doses for different scanning protocols and breast sizes were compared. Dose reduction was attempted by using partial arc CBCT scan with bowtie filter. The impact of this dose reduction strategy on image registration accuracy was investigated.
resultsThe average breast surface doses were 20.02 mGy and 11.65 mGy for thoracic CBCT without filtration and with filtration, respectively. This indicates a dose reduction of 41.8% by use of bowtie filter. It was found 220° partial arc scanning significantly reduced the dose to contralateral breast (44.4% lower than ipsilateral breast), while the image registration accuracy was not compromised.
conclusionsBreast dose reduction can be achieved by using ipsilateral 220° partial arc scan with bowtie filter. This strategy also provides sufficient image quality for thorax image registration in daily patient positioning verification.