Implementation of TomoEDGE in the independent dose calculator CheckTomo.
Article in Journal of applied clinical medical physics, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Investigation on the accuracy of field widths and the quantitative relationship with energy variations for dynamic jaws helical tomotherapy.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeCheckTomo is an independent dose calculation software for tomotherapy. Recently, Accuray (Accuray Inc., Sunnyvale, CA, USA) released an upgrade of its tomotherapy treatment device, called TomoEDGE Dynamic Jaws, which improves the quality of treatment plans by enhancing the dose delivery with the help of jaws motion. This study describes the upgrade of CheckTomo to that new feature.
methodsTo account for the varying width and off-axis shift of dynamic jaws fields, the calculation engine of CheckTomo multiplies the treatment field profile by a penumbral filter and shifts the dose calculation grid. Penumbral filters were obtained by dividing the edge field profiles by that of the corresponding nominal field. They were sampled at widths 1.0, 1.8, and 2.5 cm at isocenter in the edges of the 2.5 and 5 cm treatment field.
resultsThe upgrade of CheckTomo was tested on 30 patient treatments planned with dynamic jaws. The gamma pass rate averaged over 10 abdomen plans was 95.9%, with tolerances of 3 mm/3%. For 10 head and neck plans, the mean pass rate was 95.9% for tolerances of 4 mm/4%. Finally, misplacement and overdosage errors were simulated. In each tested cases, the 2 mm/3% gamma pass rate fell below 95% when a 4 mm shift or 3% dose difference was applied.
conclusionsThese results are equivalent to what CheckTomo achieves in static jaws cases. So, in terms of dose calculation accuracy and errors detection, the upgraded version of CheckTomo is as reliable for dynamic jaws plans as the former release was for static cases.