ArticleBMC anesthesiology2017
The value of arterial pressure waveform cardiac output measurements in the radial and femoral artery in major cardiac surgery patients.
Article in BMC anesthesiology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 1 synthesis or guideline pooled it, 18 citations in OpenAlex.
- Agreement of minimally invasive pulse wave analysis with pulmonary artery and transpulmonary thermodilution cardiac output measurements in perioperative and intensive care medicine: a systematic review and meta-analysis.British journal of anaesthesia · 2026Pooled it
- Review
- Postinduction Hypotension and Adverse Outcomes in Older Adults Undergoing Transcatheter Aortic Valve Replacement: A Retrospective Cohort Study.Clinical interventions in aging · 2024Article
- Study of the accuracy of a radial arterial pressure waveform cardiac output measurement device after cardiac surgery.Journal of cardiothoracic surgery · 2023Observational
- Non-Invasive Continuous Measurement of Haemodynamic Parameters-Clinical Utility.Journal of clinical medicine · 2021Review
- Four different methods of measuring cardiac index during cytoreductive surgery and hyperthermic intraperitoneal chemotherapy.Korean journal of anesthesiology · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundA relatively new uncalibrated arterial pressure waveform cardiac output (CO) measurement technique is the Pulsioflex-ProAQT® system. Aim of this study was to validate this system in cardiac surgery patients with a specific focus on the evaluation of a difference in the radial versus the femoral arterial access, the value of the auto-calibration modus and the ability to show fluid-induced changes.
methodsIn twenty-five patients scheduled for ascending aorta, aortic arch replacement, or both we measured CO simultaneously by transpulmonary thermodilution (COtd) and by using the ProAQT® system connected to the radial (COpR), as well as the femoral artery catheter (COpF). Hemodynamic data were assessed at predefined time points; from incision until 16 h after ICU admission.
resultsIn total 175 (radial) and 179 (femoral) pairs of CO measurement were collected. The accuracy of COpR/COpF was evaluated showing a mean bias of -0.31 L/min (±2.9 L/min) and -0.57 L/min (± 2.8 L/min) with percentage errors of 49 and 46% respectively. Trending ability of the ProAQT® device was evaluated; the four quadrant concordance rates in the radial and femoral artery were 74 and 75% and improved to 77 and 85% after auto-calibration. The mean angular biases in the radial and femoral artery were 6.4° and 6.0° and improved to 5° and 3.3° after auto-calibration. The polar concordance rates in the radial and femoral artery were 65 and 70% and improved to 76 and 84% after auto-calibration. Considering the fluid-induced changes in stroke volume(SV), the coefficient of correlation between the changes in SVtd and SVp was 0.57 (p < 0.01) in the radial artery and 0.60 (p < 0.01) in the femoral artery.
conclusionsThe ProAQT® system can be of additional value if the clinician wants to determine fluid responsiveness in cardiac surgery patients. However, the ProAQT® system provided inaccurate CO measurements compared to transpulmonary thermodilution. The trending ability was poor for COpR but moderate for COpF. Auto-calibration of the system did not improve accuracy of CO measurements nor did it improve the prediction of fluid responsiveness. However, the trending ability was improved by auto-calibration, possibly by correcting a drift over a longer time period.