Reduction of metal artifacts due to dental hardware in computed tomography angiography: assessment of the utility of model-based iterative reconstruction.

Keita Kuya, Yuki Shinohara, Ayumi Kato, Makoto Sakamoto, Masamichi Kurosaki, Toshihide Ogawa

Abstract read
PubMed Publisher
In one paragraph

Article in Neuroradiology, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 15 papers.

0numbers the graph read from it
0cells of the map it votes in
15citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

15 citing papers in PubMed.

  1. Trial
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Keita KuyaDivision of Radiology, Department of Pathophysiological and Therapeutic Science, Faculty of Medicine, Tottori University, 36-1, Nishi-cho, Yonago, 683-8504, Japan.
Yuki ShinoharaDivision of Radiology, Department of Pathophysiological and Therapeutic Science, Faculty of Medicine, Tottori University, 36-1, Nishi-cho, Yonago, 683-8504, Japan. shino-y@olive.plala.or.jp.ORCID http://orcid.org/0000-0001-6586-4086
Ayumi KatoDepartment of Radiology, Tottori Municipal Hospital, Yonago, Japan.
Makoto SakamotoDivision of Neurosurgery, Department of Neurological Sciences, Faculty of Medicine, Tottori University, Yonago, Japan.
Masamichi KurosakiDivision of Neurosurgery, Department of Neurological Sciences, Faculty of Medicine, Tottori University, Yonago, Japan.
Toshihide OgawaDivision of Radiology, Department of Pathophysiological and Therapeutic Science, Faculty of Medicine, Tottori University, 36-1, Nishi-cho, Yonago, 683-8504, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe aim of this study is to assess the value of adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR) for reduction of metal artifacts due to dental hardware in carotid CT angiography (CTA).

methodsThirty-seven patients with dental hardware who underwent carotid CTA were included. CTA was performed with a GE Discovery CT750 HD scanner and reconstructed with filtered back projection (FBP), ASIR, and MBIR. We measured the standard deviation at the cervical segment of the internal carotid artery that was affected most by dental metal artifacts (SD

resultsMBIR significantly improved AI compared with FBP and ASIR (p < 0.001, each). We found no significant difference in AI between FBP and ASIR (p = 0.502). The visual score of MBIR was significantly better than those of FBP and ASIR (p < 0.001, each), whereas the scores of ASIR were the same as those of FBP. Kappa values indicated good inter-observer agreements in all reconstructed images (0.747-0.778).

conclusionsMBIR resulted in a significant reduction in artifact from dental hardware in carotid CTA.

Indexed as

ArtifactsComputed Tomography AngiographyDental ProsthesisMetalsAgedAged, 80 and overContrast MediaFemaleHumansIohexolMaleMiddle AgedRadiographic Image Interpretation, Computer-AssistedRetrospective StudiesContrast MediaIohexolMetalsAdaptive statistical iterative reconstructionCT angiographyDental hardwareMetal artifact reductionModel-based iterative reconstruction