Dosimetry and evaluating the effect of treatment parameters on the leakage of multi leaf collimators in ONCOR linear accelerators.
Article in Advanced biomedical research, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed, 3 citations in OpenAlex.
- Comparison of Jaw Mode and Field Width for Left-Breast Cancer Using TomoDirect Three-Dimensional Conformal Radiation Therapy: A Phantom Study.Healthcare (Basel, Switzerland) · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundOne of the standard equipment in medical linear accelerators is multi-leaf collimators (MLCs); which is used as a replacement for lead shielding. MLC's advantages are a reduction of the treatment time, the simplicity of treatment, and better dose distribution. The main disadvantage of MLC is the radiation leakages from the edges and between the leaves. The purpose of this study was to determine the effect of various treatment parameters in the magnitude of MLC leakage in linear accelerators. MATERIALS AND
methodsThis project was performed with ONCOR Siemens linear accelerators. The amount of radiation leakage was determined by film dosimetry method. The films were Kodak-extended dose range-2, and the beams were 6 MV and 18 MV photons. In another part of the experiment, the fluctuation of the leakage was measured at various depths and fields.
resultsThe amount of leakage was generally up to 1.5 ± 0.2% for both energies. The results showed that the level of the leakage and the amount of dose fluctuation depends on the field size and depth of measurement. The amount of the leakage fluctuations in all energies was decreased with increasing of field size. The variation of the leakage versus field size was similar to the inverse of scattering collimator factor.
conclusionsThe amount of leakage was more for 18 MV compare to 6 MV The percentage of the leakage for both energies is less than the 5% value which is recommended by protocols. The fluctuation of the MLC leakage reduced by increasing the field size and depth.