ArticleMagnetic resonance in medicine2017
Model-based iterative reconstruction for single-shot EPI at 7T.
Article in Magnetic resonance in medicine, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Cartesian MaxGIRF: Model-based EPI reconstruction incorporating gradient nonlinearity and concomitant field effects.Magnetic resonance in medicine · 2026Article
- Blip-up blip-down circular EPI (BUDA-cEPI) for distortion-free dMRI with rapid unrolled deep learning reconstruction.Magnetic resonance imaging · 2025Article
- Reconstruction for 7T high-resolution whole-brain diffusion MRI using two-stage N/2 ghost correction and L1-SPIRiT without single-band reference.Magnetic resonance in medicine · 2023Article
- Minimizing susceptibility-induced BOLD sensitivity loss in multi-band accelerated fMRI using point spread function mapping and gradient reversal.Physics in medicine and biology · 2023Article
- Robust autocalibrated structured low-rank EPI ghost correction.Magnetic resonance in medicine · 2021Article
- Distortion-free imaging: A double encoding method (DIADEM) combined with multiband imaging for rapid distortion-free high-resolution diffusion imaging on a compact 3T with high-performance gradients.Journal of magnetic resonance imaging : JMRI · 2020Article
- Characterization of Brain Metabolism by Nuclear Magnetic Resonance.Chemphyschem : a European journal of chemical physics and physical chemistry · 2019Review
- Navigator-Free EPI Ghost Correction With Structured Low-Rank Matrix Models: New Theory and Methods.IEEE transactions on medical imaging · 2018Article
- Dynamic 2D self-phase-map Nyquist ghost correction for simultaneous multi-slice echo planar imaging.Magnetic resonance in medicine · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
Abstract
purposeTo describe a model-based reconstruction strategy for single-shot echo planar imaging (EPI) that intrinsically accounts for k-space nonuniformity, Nyquist ghosting, and geometric distortions during rather than before or after image reconstruction.
methodsRamp sampling and inhomogeneous B0 field-induced distortion cause the EPI samples to lie on a non-Cartesian grid, thus requiring the nonuniform fast Fourier transform. Additionally, a 2D Nyquist ghost phase correction without the need for extra navigator acquisition is included in the proposed reconstruction. Coil compression is also incorporated to reduce the computational load. The proposed method is applied to phantom and human brain MRI data.
resultsThe results demonstrate that Nyquist ghosting and geometric distortions are reduced by the proposed reconstruction. The proposed 2D phase correction is superior to a conventional 1D correction. The reductions of both artifacts lead to improved temporal signal-to-noise ratio (tSNR). The virtual coil results suggest that the processing time can be reduced by up to 75%, with a mean tSNR loss of only 3.2% when using 8-virtual instead of 32-physical coils for twofold undersampled data.
conclusionThe proposed reconstruction improves the quality (ghosting, geometry, and tSNR) of EPI without requiring calibration data for Nyquist ghost correction. Magn Reson Med 78:2250-2264, 2017. © 2017 International Society for Magnetic Resonance in Medicine.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials