ArticleJournal of thoracic disease2016
Cough frequency monitors: can they discriminate patient from environmental coughs?
Article in Journal of thoracic disease, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed, 1 synthesis or guideline pooled it, 17 citations in OpenAlex.
- Diagnostic Accuracy of Cough Frequency Monitors: A Systematic Review and Meta-Analysis.Respirology (Carlton, Vic.) · 2026Pooled it
- The value of cough sound monitoring via an audio-enabled smartwatch for OSA screening in COPD patients: a cross-sectional exploratory study.Frontiers in medicine · 2025Article
- An automatic cough counting method and system construction for portable devices.Frontiers in bioengineering and biotechnology · 2024Article
- Performance evaluation of human cough annotators: optimal metrics and sex differences.BMJ open respiratory research · 2023Article
- Validation of a small cough detector.ERJ open research · 2023Article
- Past and Trends in Cough Sound Acquisition, Automatic Detection and Automatic Classification: A Comparative Review.Sensors (Basel, Switzerland) · 2022Review
- Feasibility and clinical utility of ambulatory cough monitoring in an outpatient clinical setting: a real-world retrospective evaluation.ERJ open research · 2021Article
- Predictive Value of Cough Frequency in Addition to Aspiration Risk for Increased Risk of Pneumonia in Dysphagic Stroke Survivors: A Clinical Pilot Study.Brain sciences · 2021Article
- The present and future of cough counting tools.Journal of thoracic disease · 2020Review
- Article
- Treatable traits in interstitial lung disease: a narrative review.Therapeutic advances in respiratory diseaseReview
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
Abstract
backgroundObjective cough frequency measurements are increasingly applied in clinical research. Technological advances enable automated detection and counting of cough events from sound recordings of many hours' duration. A possible limitation of sound-based cough frequency measurement is the contamination of recordings by environmental coughs (coughs from persons other than the patient). This study aimed to investigate the accuracy of a sound-based cough monitor for detecting and discriminating patient cough from environmental cough.
methodsAs part of a stroke trial (ISRCTN40298220), patients on a hospital ward underwent 15-minute recordings using the Leicester Cough Monitor (LCM), a sound-based cough monitor ('semi-automated counts'). Participants and other persons in the environment were prompted to cough. An observer present in the room recorded the number of patient and environmental coughs ('live counts'). LCM counts were also compared against a manual cough count, the most commonly used gold standard to determine accuracy ('manual sound counts' from listening to recordings), by a blinded assessor who cross-referenced timed cough events from the respective methods. Data for automated, manual and live cough counts were analyzed using agreement statistics.
resultsOn sound recordings from five patients, there were 65 patient coughs and 78 environmental coughs (manual counts). Absolute agreement for patient cough count between all three measurement methods (LCM automated, live, and manual sound counts) was high, with intra-class correlation coefficient of 0.94 [95% confidence intervals (CI): 0.74, 0.99]. The proportion of exact agreements for patient cough between LCM and manual count was 0.92, and kappa was 0.84 (95% CI: 0.75, 0.93). The LCM showed sensitivity of 0.94 (95% CI: 0.84, 0.98), specificity of 0.91 (95% CI: 0.82, 0.96), positive predictive value of 0.90 (95% CI: 0.79, 0.95) and negative predictive value of 0.95 (95% CI: 0.86, 0.98) for detecting patient coughs.
conclusionsThis preliminary study supports the validity of the cough monitor for detecting and discriminating patient from environmental cough. Further validation is recommended, to describe the level of accuracy with greater precision.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.