ArticleThe journal of advanced prosthodontics2016
Influence of the connection design and titanium grades of the implant complex on resistance under static loading.
Article in The journal of advanced prosthodontics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed.
- Optimizing dental implant design: Structure, strength, and bone ingrowth.Journal of dental sciences · 2025Article
- Biofilm Removal from In Vitro Narrow Geometries Using Single and Dual Pulse Er:YAG Laser Photoacoustic Irrigation.Microorganisms · 2023Article
- Shear Bond Strength of Lithium Disilicate Bonded with Various Surface-Treated Titanium.International journal of dentistry · 2022Article
- Descriptive retrospective study analyzing relevant factors related to dental implant failure.Medicina oral, patologia oral y cirugia bucal · 2019Article
- Improvement in Fatigue Behavior of Dental Implant Fixtures by Changing Internal Connection Design: An In Vitro Pilot Study.Materials (Basel, Switzerland) · 2019Article
- Deformation of the Internal Connection of Narrow Implants after Insertion in Dense Bone: An in Vitro Study.Materials (Basel, Switzerland) · 2019Article
- Complete mechanical characterization of an external hexagonal implant connection: in vitro study, 3D FEM, and probabilistic fatigue.Medical & biological engineering & computing · 2018Article
- Load-Bearing Capacity and Retention of Newly Developed Micro-Locking Implant Prosthetic System: An In Vitro Pilot Study.Materials (Basel, Switzerland) · 2018Article
- A New Model to Study Fatigue in Dental Implants Based on Probabilistic Finite Elements and Cumulative Damage Model.Applied bionics and biomechanics · 2017Article
- Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections.Dental research journalArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe purpose of this study was to evaluate the resistance to deformation under static overloading by measuring yield and fracture strength, and to analyze the failure characteristics of implant assemblies made of different titanium grades and connections. MATERIALS AND
methodsSix groups of implant assemblies were fabricated according to ISO 14801 (n=10). These consisted of the combinations of 3 platform connections (external, internal, and morse tapered) and 2 materials (titanium grade 2 and titanium grade 4). Yield strength and fracture strength were evaluated with a computer-controlled Universal Testing Machine, and failed implant assemblies were classified and analyzed by optical microscopy. The data were analyzed using the One-way analysis of variance (ANOVA) and Student's t-test with the level of significance at
resultsThe group IT4
conclusionA longer internal connection and titanium grade 4 of the implant system is advantageous for static overloading condition. However, it is not only the connection design that affects the stability. The strength of the titanium grade as material is also important since it affects the implant stability. When using the implant system made of titanium grade 2, a larger diameter fixture should be selected in order to provide enough strength to withstand overloading.