Local filtration based scatter correction for cone-beam CT using primary modulation.
Article in Medical physics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- Evaluation of the feasibility of a multisource CBCT for maxillofacial imaging.Physics in medicine and biology · 2023Article
- Development and validation of a scatter-corrected CBCT image-guided method for cervical cancer brachytherapy.Frontiers in oncology · 2022Article
- Fast shading correction for cone-beam CT via partitioned tissue classification.Physics in medicine and biology · 2019Article
- Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT.Technology in cancer research & treatment · 2019Article
- 4D cone-beam computed tomography (CBCT) using a moving blocker for simultaneous radiation dose reduction and scatter correction.Physics in medicine and biology · 2018Article
- The role of off-focus radiation in scatter correction for dedicated cone beam breast CT.Medical physics · 2018Article
- Optimization of the geometry and speed of a moving blocker system for cone-beam computed tomography scatter correction.Medical physics · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
1 author.
Funding
Abstract
purposeExcessive scatter contamination fundamentally limits the image quality of cone-beam CT (CBCT), hindering its quantitative use in clinical applications. The author has previously proposed an effective scatter correction method for CBCT using primary modulation. A Fourier transform-based algorithm (FTPM) was implemented to estimate scatter from modulated projections, with a few limitations including the assumption of uniform modulation frequency and magnitude that becomes less accurate in the presence of beam-hardening and other nonideal effects. This paper aims to overcome the above drawbacks by developing a new algorithm for the primary modulation method with improved accuracy and reliability.
methodsIncident x-ray intensities for each detector pixel with and without the interception of the modulator blocker are estimated from a modulated flat-field image. A new signal relationship is then developed to obtain a first scatter estimate from a modulated projection using a spatially varying modulation distribution. The method empirically adjusts the effective modulation magnitude for each projection ray to account for the beam-hardening effects. Estimated scatter signals with high expected errors are discarded in the generation of the final scatter distribution. The author proposes a technique of local filtration to accelerate major portions of the signal processing, and the new algorithm is referred to as local filtration based primary modulation (LFPM).
resultsThe study on the Catphan® 600 phantom shows that LFPM effectively removes scatter-induced cupping artifacts on CBCT images and reduces the CT image error from 222 to 15 HU. In addition, the image contrast on eight contrast rods of the phantom is enhanced by a factor of 2 on average. On an anthropomorphic head phantom, LFPM reduces the CT image error from 153 to 18 HU and eliminates the streak artifacts observed on the result of FTPM with substantially improved image uniformity. On the Rando® phantom, LFPM reduces the CT image error from 278 to 4 HU around the object center.
conclusionsAs compared with the previously developed FTPM algorithm, LFPM enhances the imaging performance by using a more flexible data processing framework that does not require projection data downsampling or uniform modulation frequency and magnitude. It also becomes possible to discard suspicious scatter estimate values prior to the generation of a final scatter distribution and to model the beam-hardening effects on modulation for improved scatter estimation accuracy. The presented research further exploits the potential of the primary modulation method on scatter correction and facilitates its clinical adoption in CBCT imaging.