ArticleNeuromodulation : journal of the International Neuromodulation Society2016
Intrathecal Drug Delivery Systems (IDDS): The Implantable Systems Performance Registry (ISPR).
Article in Neuromodulation : journal of the International Neuromodulation Society, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to 2 registered trials, which are not on this map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Transitioning to a Valve -Gated Intrathecal Drug Delivery System (IDDS): An Open-Label, Non-Randomized, Single-Blind, Multi-Center, Cross-Over Study (TRANSIT)
Transitioning to a Valve -Gated Intrathecal Drug Delivery System (IDDS): An Open-Label, Non-Randomized, Single-Blind, Multi-Center Study
Who cites it
11 citing papers in PubMed.
- Intrathecal Drug Delivery Systems for Cancer Pain: An Analysis of a Prospective, Multicenter Product Surveillance Registry.Anesthesia and analgesia · 2020Trial
- Article
- Catheter dysfunction in long-term intrathecal baclofen pump users and when to evaluate integrity: A case series.Interventional pain medicine · 2024Article
- Position-Dependent Intrathecal Baclofen System Catheter Failure Resulting in Debilitating Spasticity: A Case Report.Cureus · 2024Article
- Interventional Pain Procedures: A Narrative Review Focusing On Safety and Complications. PART 2 Interventional Procedures For Back Pain.Journal of pain research · 2023Review
- Intrathecal Drug Delivery: Advances and Applications in the Management of Chronic Pain Patient.Frontiers in pain research (Lausanne, Switzerland) · 2022Review
- Adverse Events and Complications Associated With Intrathecal Drug Delivery Systems: Insights From the Manufacturer and User Facility Device Experience (MAUDE) Database.Neuromodulation : journal of the International Neuromodulation Society · 2021Article
- Targeted Drug Delivery for Chronic Nonmalignant Pain: Longitudinal Data From the Product Surveillance Registry.Neuromodulation : journal of the International Neuromodulation Society · 2021Article
- Intrathecal Baclofen for Severe Spasticity: Longitudinal Data From the Product Surveillance Registry.Neuromodulation : journal of the International Neuromodulation Society · 2020Article
- Occlusion of the Ascenda Catheter in a Patient with Pump Twiddler's Sydrome: A Case Report.Anesthesiology and pain medicine · 2018Article
- Retrospective analysis of the financial break-even point for intrathecal morphine pump use in Korea.The Korean journal of pain · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesThe ISPR was initially created to monitor the product performance of Medtronic implanted intrathecal drug infusion and spinal cord systems available in the United States. MATERIALS AND
methodsData were collected from 50 representative sites implanting and following patients with intrathecal drug delivery systems across the United States between August 7, 2003 and January 31, 2014. Device performance over time was estimated using life table survival methods.
resultsOf the 6093 patients enrolled in the ISPR, 3405 (55.9%) were female and 2675 (43.9%) were male, and 13 (0.2%) did not provide gender data. The average age at enrollment was 52.9 years (SD =17.6 years) and average follow-up time was 29.6 months. Currently, the estimates of device survival from pump-related events exceed 90% for all pump models across the applicable follow-up time points. The majority of product performance events were catheter-related. At 5 years of follow-up, all applicable catheter models, with the exception of revised not as designed or grafted not as designed catheters, had greater than 81% survival from catheter-related events.
conclusionsThe ISPR is designed to serve as an ongoing source of system and device-related information with a focus on "real-world" safety and product performance. ISPR data continue to be used to guide future product development efforts aimed at improving product reliability and quality.