Evaluation of Doses and Image Quality in Mammography with Screen-Film, CR, and DR Detectors - Application of the ACR Phantom.

Wioletta Ślusarczyk-Kacprzyk, Witold Skrzyński, Ewa Fabiszewska

Abstract read
In one paragraph

Article in Polish journal of radiology, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Characterization of the imaging settings in screening mammography using a tracking and reporting system: A multi-center and multi-vendor analysis.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2020
    Article
  3. Article
  4. Article
  5. Article
  6. Investigation of Exposure Factors for Various Breast Composition and Thicknesses in Digital Screening Mammography Related to Breast Dose.Medical principles and practice : international journal of the Kuwait University, Health Science Centre · 2018
    Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Wioletta Ślusarczyk-KacprzykDepartment of Medical Physics, Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw, Poland.
Witold SkrzyńskiDepartment of Medical Physics, Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw, Poland.
Ewa FabiszewskaDepartment of Medical Physics, Maria Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw, Poland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundDifferent methods of image quality evaluation are routinely used for analogue and digital mammography systems in Poland. In the present study, image quality for several screen-film (SF), computed radiography (CR), and fully digital (DR) mammography systems was compared directly with the use of the ACR mammography accreditation phantom. MATERIAL/

methodsImage quality and mean glandular doses were measured and compared for 47 mammography systems in the Mazovia Voivodeship in Poland, including 26 SF systems, 12 CR systems, and 9 DR systems. The mean glandular dose for the breast simulated by 4.5 cm of PMMA was calculated with methods described in the "European guidelines for quality assurance in breast cancer screening and diagnosis". Visibility of the structures in the image (fibers, microcalcifications, and masses) was evaluated with the mammographic accreditation ACR phantom.

resultsImage quality for DR systems was significantly higher than for SF and CR systems. Several SF systems failed to pass the image quality tests because of artifacts. The doses were within acceptable limits for all of the systems, but the doses for the CR systems were significantly higher than for the SF and DR systems.

conclusionsThe best image quality, at a reasonably low dose, was observed for the DR systems. The CR systems are capable of obtaining the same image quality as the SF systems, but only at a significantly higher dose. The ACR phantom can be routinely used to evaluate image quality for all types of mammographic systems.

Indexed as

MammographyQuality ControlRadiation Dosage

Identifiers

PMID27617048
PMCPMC4993217