Evidence map›Paper›PMID 27604905›Full record

ArticleEuropean journal of orthopaedic surgery & traumatology : orthopedie traumatologie2017

Verification of in vivo accuracy of Trumatch™ patient-specific instrumentation in total knee replacement using pin-less computer navigation.

Nikolai Briffa, Mohamed A Imam, Ravi Mallina, Ashraf Abdelkafy, Ajeya Adhikari

Abstract readEvaluation Study
PubMed Publisher
In one paragraph

Article in European journal of orthopaedic surgery & traumatology : orthopedie traumatologie, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Nikolai BriffaThe South West London Elective Orthopaedic Centre, Dorking Road, Epsom, London, KT18 7EG, UK.
Mohamed A ImamThe South West London Elective Orthopaedic Centre, Dorking Road, Epsom, London, KT18 7EG, UK.
Ravi MallinaThe South West London Elective Orthopaedic Centre, Dorking Road, Epsom, London, KT18 7EG, UK.
Ashraf AbdelkafyOrthopaedic Surgery and Traumatology Department, Faculty of Medicine, Suez Canal University, Circular Road, Ismailia, 41522, Egypt. ashraf.abdelkafy@gmail.com.
Ajeya AdhikariThe South West London Elective Orthopaedic Centre, Dorking Road, Epsom, London, KT18 7EG, UK.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeAccurate component alignment in total knee replacement (TKR) is one of the important factors in determining long-term survivorship. This has been achieved by conventional jigs, computer-assisted technology (CAS) and more recently patient-specific instrumentation (PSI). The purpose of the current study was to investigate the in vivo accuracy of Trumatch™ PSI using validated pin-less computer navigation system.

methodTwenty consecutive selected patients that fulfilled our inclusion/exclusion criteria underwent TKR using PSI. Coronal alignment, posterior slope, resection thickness and femoral sagittal alignment were recorded using pin-less navigation. The position of the actual cutting block was appropriately adjusted prior to proceeding to definitive resections.

resultsThe coronal alignment using PSI without the assistance of navigation would have resulted in 14 (70 %) within ±3°, 11 (55 %) within ±2° and 6 (30 %) outside acceptable alignment. Thirty-five percentage of proposed femur sagittal alignment and 55 % of posterior tibial slope were achieved within ±3°. Components size was accurately predicted in 95 % of femurs and 90 % of tibia.

conclusionThe purported advantages in restoring alignments using Trumatch™ PSI alone over standard equipment are debatable. However, it predicts sizing well, and femoral coronal alignment is reasonable. Combining Trumatch™ PSI with CAS will allow in vivo verification and necessary corrections. LEVEL OF EVIDENCE: IV.

Indexed as

AgedArthroplasty, Replacement, KneeBone MalalignmentFemaleFemurHumansMaleOsteoarthritis, KneePatient Care PlanningPreoperative CareProspective StudiesSurgery, Computer-AssistedTibiaComputer navigationPatient-specific instrumentationTotal knee arthroplastyTotal knee replacementTrumatch™

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials