Evidence map›Paper›PMID 27603720›Full record

ArticleInternational journal for numerical methods in biomedical engineering2017

Computational virtual evaluation of the effect of annuloplasty ring shape.

Ahnryul Choi, David D McPherson, Hyunggun Kim

Abstract read
In one paragraph

Article in International journal for numerical methods in biomedical engineering, 2017. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
0.4field-weighted citation impact, top 29% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 13 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 2 countries.

Ahnryul ChoiDivision of Cardiovascular Medicine Department of Internal Medicine, The University of Texas Health Science Center at Houston, Houston, TX, USA.
David D McPhersonDivision of Cardiovascular Medicine Department of Internal Medicine, The University of Texas Health Science Center at Houston, Houston, TX, USA.
Hyunggun KimDivision of Cardiovascular Medicine Department of Internal Medicine, The University of Texas Health Science Center at Houston, Houston, TX, USA.
Sungkyunkwan University · KRThe University of Texas Health Science Center at Houston · US

Funding

Mitral Valve Dynamic Analysis And Potential Clinical ApplicationsR01HL109597 · NHLBI · UNIVERSITY OF TEXAS HLTH SCI CTR HOUSTON · PI MCPHERSON, DAVID D · 2011 to 2015
$1.8M
NHLBI NIH HHS R01 HL109597
6 · The paper itself

Abstract

Mitral regurgitation (MR) is a result of mitral valve (MV) pathology. Its etiology can be categorized as degenerative or functional MR. Ring annuloplasty aims to reconfigure a dilated mitral annulus to its normal size and shape. We investigated the effect of annuloplasty ring shape on MR outcome using our established 3-dimensional (3-D) echocardiography-based computational MV evaluation protocols. Virtual patient MV models were created from 3-D transesophageal echocardiographic data in patients with MR because of mitral annular dilation. Two distinct annuloplasty rings (Physio II and GeoForm) were designed and virtually implanted to the patient MVs. Dynamic finite element simulations of MV function were performed for each MV after virtual ring annuloplasty of either ring, and physiologic and biomechanical characteristics of MV function were compared. Excessive stress values appeared primarily in the midanterior and midposterior regions, and lack of leaflet coaptation was found in pre-annuloplasty patient MVs. Both rings demonstrated marked reduction of stresses and efficient leaflet coaptation. The Physio II ring demonstrated more evenly distributed stress reduction across the leaflets and annulus compared with the GeoForm ring. Conversely, the highly nonplanar curvature of the GeoForm ring more effectively increased leaflet coaptation compared with the Physio II ring. This indicates that the shape of annuloplasty ring affects post-annuloplasty physiologic and biomechanical conditions, which can lead to tissue alteration over a longer period after ring annuloplasty. This virtual ring annuloplasty simulation strategy provides detailed physiologic and biomechanical information and may help better plan the optimal ring selection and improved patient-specific MV repairs.

Indexed as

Heart Valve ProsthesisEchocardiography, Three-DimensionalHumansMitral ValveMitral Valve InsufficiencyTreatment Outcomecomputational simulationmitral regurgitationmitral valvemitral valve repairring annuloplastyvirtual surgery

Identifiers

PMID27603720
PMCPMC5340636
OpenAlexW2511338411

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.