Multisource inverse-geometry CT. Part II. X-ray source design and prototype.
Article in Medical physics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 23 citations in OpenAlex.
- Overcoming the thermal limits of photon counting CT resolution using focal spot multiplexing: A feasibility study.Medical physics · 2025Article
- Evaluation of the feasibility of a multisource CBCT for maxillofacial imaging.Physics in medicine and biology · 2023Article
- Volumetric computed tomography with carbon nanotube X-ray source array for improved image quality and accuracy.Communications engineering · 2023Article
- Design and optimization of thin-film tungsten (W)-diamond target for multi-pixel X-ray sources.Medical physics · 2022Article
- A simulated comparison of lung tumor target verification using stereoscopic tomosynthesis or radiography.Medical physics · 2022Article
- Tetrahedron beam imaging with a multi-pixel thermionic emission X-ray source and a photon-counting detector: A benchtop experimental study.Medical physics · 2021Article
- A prototype Multi-X-ray-source array (MXA) for digital breast tomosynthesis.Physics in medicine and biology · 2020Article
- Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations.Physics in medicine and biology · 2020Article
- Multisource inverse-geometry CT. Part I. System concept and development.Medical physics · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
15 authors at 2 institutions in 1 country.
Funding
Abstract
purposeThis paper summarizes the development of a high-power distributed x-ray source, or "multisource," designed for inverse-geometry computed tomography (CT) applications [see B. De Man et al., "Multisource inverse-geometry CT. Part I. System concept and development," Med. Phys. 43, 4607-4616 (2016)]. The paper presents the evolution of the source architecture, component design (anode, emitter, beam optics, control electronics, high voltage insulator), and experimental validation.
methodsDispenser cathode emitters were chosen as electron sources. A modular design was adopted, with eight electron emitters (two rows of four emitters) per module, wherein tungsten targets were brazed onto copper anode blocks-one anode block per module. A specialized ceramic connector provided high voltage standoff capability and cooling oil flow to the anode. A matrix topology and low-noise electronic controls provided switching of the emitters.
resultsFour modules (32 x-ray sources in two rows of 16) have been successfully integrated into a single vacuum vessel and operated on an inverse-geometry computed tomography system. Dispenser cathodes provided high beam current (>1000 mA) in pulse mode, and the electrostatic lenses focused the current beam to a small optical focal spot size (0.5 × 1.4 mm). Controlled emitter grid voltage allowed the beam current to be varied for each source, providing the ability to modulate beam current across the fan of the x-ray beam, denoted as a virtual bowtie filter. The custom designed controls achieved x-ray source switching in <1 μs. The cathode-grounded source was operated successfully up to 120 kV.
conclusionsA high-power, distributed x-ray source for inverse-geometry CT applications was successfully designed, fabricated, and operated. Future embodiments may increase the number of spots and utilize fast read out detectors to increase the x-ray flux magnitude further, while still staying within the stationary target inherent thermal limitations.