Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems.
Article in PloS one, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed.
- Intra-vendor inter-scanner agreement of negative-HU measurements on fast kV-switching dual-energy CT virtual monoenergetic imaging: a phantom study.Radiological physics and technology · 2026Article
- Impact of differences in computed tomography value-electron density/physical density conversion tables on calculate dose in low-density areas.Physical and engineering sciences in medicine · 2025Article
- Article
- A Dosimetric Comparison Study for Blood Irradiation Employing Different Medium and Algorithms in Clinical Linear Accelerator.Journal of blood medicine · 2024Article
- Radiation dose and cancer risks from radiation exposure during abdominopelvic computed tomography (CT) scans: comparison of diagnostic and radiotherapy treatment planning CT scans.Radiation and environmental biophysics · 2021Article
- The impact of mass density variations on an electron Monte Carlo algorithm for radiotherapy dose calculations.Physics and imaging in radiation oncology · 2018Article
- Dosimetric impact of tumor treating field (TTField) transducer arrays onto treatment plans for glioblastomas - a planning study.Radiation oncology (London, England) · 2018Article
- Liquid tissue surrogates for X-ray and CT phantom studies.Medical physics · 2017Article
- Investigation of the Effect of Calibration Curves Obtained from Different Computed Tomography Devices on the Dose Distribution of Tomotherapy Plans.Journal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionFor accurate dose calculations, it is necessary to provide a correct relationship between the CT numbers and electron density in radiotherapy treatment planning systems (TPSs). The purpose of this study was to investigate the energy dependence of measured CT numbers on substituted materials used for CT number calibration of radiotherapy TPSs and the resulting errors in the treatment planning calculation doses. MATERIALS AND
methodsIn this study, we designed a cylindrical water phantom with different materials used as tissue equivalent materials for the simulation of tissues and obtaining the related CT numbers. For evaluating the effect of CT number variations of substituted materials due to energy changing of scanner (kVp) on the dose calculation of TPS, the slices of the scanned phantom at three kVp's were imported into the desired TPSs (MIRS and CorePLAN). Dose calculations were performed on two TPSs.
resultsThe mean absolute percentage differences between the CT numbers of CT scanner and two treatment planning systems for all the samples were 3.22%±2.57% for CorePLAN and 2.88%±2.11% for MIRS. It was also found that the maximum absolute percentage difference between all of the calculated doses from each photon beam of linac (6 and 15 MV) at three kVp's was less than 1.2%. DISCUSSION: The present study revealed that, for the materials with effective low atomic number, the mean CT number increased with increasing energy, which was opposite for the materials with an effective high atomic number. We concluded that the tissue substitute materials had a different behavior in the energy ranges from 80 to 130 kVp. So, it is necessary to consider the energy dependence of the substitute materials used for the measurement or calibration of CT number for radiotherapy treatment planning systems.