ArticleJournal of Zhejiang University. Science. B2016
Femoral indicator injection for transpulmonary thermodilution using the EV1000/VolumeView(®): do the same criteria apply as for the PiCCO(®)?
Article in Journal of Zhejiang University. Science. B, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 12 citations in OpenAlex.
- Accuracy of hemodynamic parameters derived by GE E-PiCCO in comparison with PiCCO® in patients admitted to the intensive care unit.Scientific reports · 2023Article
- Comparison of global end-diastolic volume index derived from jugular and femoral indicator injection: a prospective observational study in patients equipped with both a PiCCO-2 and an EV-1000-device.Scientific reports · 2020Observational
- Aortic volume determines global end-diastolic volume measured by transpulmonary thermodilution.Intensive care medicine experimental · 2020Article
- Mandatory criteria for the application of variability-based parameters of fluid responsiveness: a prospective study in different groups of ICU patients.Journal of Zhejiang University. Science. B · 2018Article
- Observational
- Applicability of stroke volume variation in patients of a general intensive care unit: a longitudinal observational study.Journal of clinical monitoring and computing · 2017Observational
- Transpulmonary thermodilution: advantages and limits.Critical care (London, England) · 2017Review
- Observational
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveComparison of global end-diastolic volume index (GEDVI) obtained by femoral and jugular transpulmonary thermodilution (TPTD) indicator injections using the EV1000/VolumnView(®) device (Edwards Lifesciences, Irvine, USA).
methodsIn an 87-year-old woman with hypovolemic shock and equipped with both jugular and femoral vein access and monitored with the EV1000/VolumeView(®) device, we recorded 10 datasets, each comprising duplicate TPTD via femoral access and duplicate TPTD (20 ml cold saline) via jugular access.
resultsMean femoral GEDVI ((674.6±52.3) ml/m(2)) was significantly higher than jugular GEDVI ((552.3±69.7) ml/m(2)), with P=0.003. Bland-Altman analysis demonstrated a bias of (+122±61) ml/m(2), limits of agreement of -16 and +260 ml/m(2), and a percentage error of 22%. Use of the correction-formula recently suggested for the PiCCO(®) device significantly reduced bias and percentage error. Similarly, mean values of parameters derived from GEDVI such as pulmonary vascular permeability index (PVPI; 1.244±0.101 vs. 1.522±0.139; P<0.001) and global ejection fraction (GEF; (24.7±1.6)% vs. (28.1±1.8)%; P<0.001) were significantly different in the case of femoral compared to jugular indicator injection. Furthermore, the mean cardiac index derived from femoral indicator injection ((4.50±0.36) L/(min·m²)) was significantly higher (P=0.02) than that derived from jugular indicator injection ((4.12±0.44) L/(min·m²)), resulting in a bias of (+0.38±0.37) L/(min·m²) and a percentage error of 19.4%.
conclusionsFemoral access for indicator injection results in markedly altered values provided by the EV1000/VolumeView(®), particularly for GEDVI, PVPI, and GEF.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials