Development of a stent-biovalve with round-shaped leaflets: in vitro hydrodynamic evaluation for transcatheter pulmonary valve implantation (TPVI).

Hirohito Sumikura, Yasuhide Nakayama, Kentaro Ohnuma, Yoshiaki Takewa, Eisuke Tatsumi

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Development of transcatheter implantable autologous tissue-engineered pulmonary valves using in-body tissue architecture.Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs · 2025
    Article
  2. Review
  3. Article
  4. Journal of Artificial Organs 2016: the year in review : Journal of Artificial Organs Editorial Committee.Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs · 2017
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Hirohito SumikuraDepartment of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita-Shi, Osaka, 565-8565, Japan. sumikura@ri.ncvc.go.jp.
Yasuhide NakayamaDivision of Medical Engineering and Materials, National Cerebral and Cardiovascular Center Research Institute, Suita-Shi, Japan.
Kentaro OhnumaDepartment of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita-Shi, Osaka, 565-8565, Japan.
Yoshiaki TakewaDepartment of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita-Shi, Osaka, 565-8565, Japan.
Eisuke TatsumiDepartment of Artificial Organs, National Cerebral and Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita-Shi, Osaka, 565-8565, Japan.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

This study evaluates a newly designed autologous heart valve-shaped tissue with a stent [stent-biovalve (SBV)] for transcatheter pulmonary valve implantation using the "in-body tissue architecture" technology. In the previously developed SBV with flat-shaped leaflets (FS-SBV), the valve could not close rapidly, because the leaflets were fixed in the open position, which induced regurgitant volume in the closing phase. Therefore, a novel mold to fabricate an SBV with round-shaped leaflets (RS-SBV) was developed, and its hydrodynamic performance with different valve diameters was evaluated in this study. A specially designed, self-expandable, stent-mounted, acrylic mold, which has 3 hemispheres, was placed in dorsal subcutaneous pouches of goats for 2 months. After extraction, the acrylic mold was removed from the implant, and a tubular tissue impregnated with the stent strut was obtained. Half of the tubular tissue with 3 hemispheres was completely folded in half inwards. The acrylic mold was designed, such that the folded half of the tubular tissue became the round-shaped leaflets. The 3 commissure parts were connected to form 3 leaflets, resulting in the preparation of the RS-SBV (internal diameter 25 mm). The RS-SBV closed more rapidly than the FS-SBV in a pulsatile mock circulation circuit under the pulmonary circulation conditions. The regurgitant fraction of the RS-SBV was approximately 6 %, which was lower than that of the FS-SBV. The appropriate pulmonary annulus diameter of the RS-SBV was from 24 to 25 mm based on the pressure difference and effective orifice area.