A controlled statistical study to assess measurement variability as a function of test object position and configuration for automated surveillance in a multicenter longitudinal COPD study (SPIROMICS).
Article in Medical physics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT01969344 (Subpopulations and Intermediate Outcome Measures in COPD Study), which is not on this map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Subpopulations and Intermediate Outcome Measures in COPD Study
Who cites it
6 citing papers in PubMed.
- Comparison of quantitative lung measures in low dose energy-integrating detector and photon-counting detector chest CT with an anthropomorphic phantom.Biomedical physics & engineering express · 2025Article
- Design of the SPIROMICS Study of Early COPD Progression: SOURCE Study.Chronic obstructive pulmonary diseases (Miami, Fla.) · 2024Article
- Filtration-histogram based texture analysis and CALIPER based pattern analysis as quantitative CT techniques in idiopathic pulmonary fibrosis: head-to-head comparison.The British journal of radiology · 2022Article
- Radiographic lung volumes predict progression to COPD in smokers with preserved spirometry in SPIROMICS.The European respiratory journal · 2019Observational
- Validated imaging biomarkers as decision-making tools in clinical trials and routine practice: current status and recommendations from the EIBALL* subcommittee of the European Society of Radiology (ESR).Insights into imaging · 2019Article
- SPIROMICS Protocol for Multicenter Quantitative Computed Tomography to Phenotype the Lungs.American journal of respiratory and critical care medicine · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
purposeA test object (phantom) is an important tool to evaluate comparability and stability of CT scanners used in multicenter and longitudinal studies. However, there are many sources of error that can interfere with the test object-derived quantitative measurements. Here the authors investigated three major possible sources of operator error in the use of a test object employed to assess pulmonary density-related as well as airway-related metrics.
methodsTwo kinds of experiments were carried out to assess measurement variability caused by imperfect scanning status. The first one consisted of three experiments. A COPDGene test object was scanned using a dual source multidetector computed tomographic scanner (Siemens Somatom Flash) with the Subpopulations and Intermediate Outcome Measures in COPD Study (SPIROMICS) inspiration protocol (120 kV, 110 mAs, pitch = 1, slice thickness = 0.75 mm, slice spacing = 0.5 mm) to evaluate the effects of tilt angle, water bottle offset, and air bubble size. After analysis of these results, a guideline was reached in order to achieve more reliable results for this test object. Next the authors applied the above findings to 2272 test object scans collected over 4 years as part of the SPIROMICS study. The authors compared changes of the data consistency before and after excluding the scans that failed to pass the guideline.
resultsThis study established the following limits for the test object: tilt index ≤0.3, water bottle offset limits of [-6.6 mm, 7.4 mm], and no air bubble within the water bottle, where tilt index is a measure incorporating two tilt angles around x- and y-axis. With 95% confidence, the density measurement variation for all five interested materials in the test object (acrylic, water, lung, inside air, and outside air) resulting from all three error sources can be limited to ±0.9 HU (summed in quadrature), when all the requirements are satisfied. The authors applied these criteria to 2272 SPIROMICS scans and demonstrated a significant reduction in measurement variation associated with the test object.
conclusionsThree operator errors were identified which significantly affected the usability of the acquired scan images of the test object used for monitoring scanner stability in a multicenter study. The authors' results demonstrated that at the time of test object scan receipt at a radiology core laboratory, quality control procedures should include an assessment of tilt index, water bottle offset, and air bubble size within the water bottle. Application of this methodology to 2272 SPIROMICS scans indicated that their findings were not limited to the scanner make and model used for the initial test but was generalizable to both Siemens and GE scanners which comprise the scanner types used within the SPIROMICS study.