Evidence map›Paper›PMID 27094510›Full record

ArticleBMC medical imaging2016

3D rotational fluoroscopy for intraoperative clip control in patients with intracranial aneurysms--assessment of feasibility and image quality.

Thomas Westermaier, Thomas Linsenmann, György A Homola, Mario Loehr, Christian Stetter, Nadine Willner, Ralf-Ingo Ernestus, Laszlo Solymosi, Giles H Vince

Open access · goldAbstract read
In one paragraph

Article in BMC medical imaging, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
1.1field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 10 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. [Value of 640-slice 3D CT angiography plus 3D printing for improving surgeries for intracranial aneurysms].Nan fang yi ke da xue xue bao = Journal of Southern Medical University · 2017
    Article
  5. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 2 institutions in 2 countries.

Thomas WestermaierDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany. Westermaie_T@ukw.de.
Thomas LinsenmannDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
György A HomolaDepartment of Neuroradiology, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Mario LoehrDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Christian StetterDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Nadine WillnerDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Ralf-Ingo ErnestusDepartment of Neurosurgery, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Laszlo SolymosiDepartment of Neuroradiology, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany.
Giles H VinceAbteilung für Neurochirurgie, Klinikum Klagenfurt, Feschnigstraße 11, 9020, Klagenfurt am Woerthersee, Austria.
University of Würzburg · DEKlinikum Klagenfurt · AT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundMobile 3D fluoroscopes have become increasingly available in neurosurgical operating rooms. In this series, the image quality and value of intraoperative 3D fluoroscopy with intravenous contrast agent for the evaluation of aneurysm occlusion and vessel patency after clip placement was assessed in patients who underwent surgery for intracranial aneurysms. MATERIALS AND

methodsTwelve patients were included in this retrospective analysis. Prior to surgery, a 360° rotational fluoroscopy scan was performed without contrast agent followed by another scan with 50 ml of intravenous iodine contrast agent. The image files of both scans were transferred to an Apple PowerMac® workstation, subtracted and reconstructed using OsiriX® free software. The procedure was repeated after clip placement. Both image sets were compared for assessment of aneurysm occlusion and vessel patency.

resultsImage acquisition and contrast administration caused no adverse effects. Image quality was sufficient to follow the patency of the vessels distal to the clip. Metal artifacts reduce the assessability of the immediate vicinity of the clip. Precise image subtraction and post-processing can reduce metal artifacts and make the clip-site assessable and depict larger neck-remnants.

conclusionThis technique quickly supplies images at adequate quality to evaluate distal vessel patency after aneurysm clipping. Significant aneurysm remnants may be depicted as well. As it does not require visual control of all vessels that are supposed to be evaluated intraoperatively, this technique may be complementary to other intraoperative tools like indocyanine green videoangiography and micro-Doppler, especially for the assessment of larger aneurysms. At the momentary state of this technology, it cannot replace postoperative conventional angiography. However, 3D fluoroscopy and image post-processing are young technologies. Further technical developments are likely to result in improved image quality.

Indexed as

Contrast MediaFeasibility StudiesFluoroscopyHumansImage Interpretation, Computer-AssistedImaging, Three-DimensionalIntracranial AneurysmNeurosurgical ProceduresRetrospective StudiesSoftwareContrast Media3D fluoroscopyAneurysm surgeryAngiographyClip controlContrastImage qualityIntraoperativePost-processingVessel patency

Identifiers

PMID27094510
PMCPMC4837534
OpenAlexW2336568452

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.