ArticleJournal of medical radiation sciences2016
Evaluation of accuracy and reproducibility of a relocatable maxillary fixation system for fractionated intracranial stereotactic radiation therapy.
Article in Journal of medical radiation sciences, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 6 citations in OpenAlex.
- Positioning error analysis of the fraxion localization system in the intracranial stereotactic radiotherapy of tumors.Clinical & translational oncology : official publication of the Federation of Spanish Oncology Societies and of the National Cancer Institute of Mexico · 2021Article
- Setup errors in radiation therapy for thoracic tumor patients of different body mass index.Journal of applied clinical medical physics · 2018Article
- Evaluation of accuracy and reproducibility of a relocatable maxillary fixation system for fractionated intracranial stereotactic radiation therapy.Journal of medical radiation sciences · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionAccurate localisation is an essential component for the delivery of intracranial stereotactic treatment. For fractionated stereotactic radiotherapy, we compared the daily localisation accuracy of a standard thermoplastic mask with a new maxillary fixation device (MFD).
methodsDaily pre-treatment kV cone-beam computed tomography (CBCT) scans of 23 patients (12 localised in the MFD and 11 in the mask) with benign skull-based lesions were reviewed retrospectively. The set up accuracy was measured in 6° of freedom, to ascertain both individual and population random and systematic errors. The appropriate clinical target volume to planning target volume margin was computed from set up error data.
resultsA total of 682 CBCT scans were evaluated. Systematic (Σ) and random (σ) population errors were Σ = 0.8 mm, 0.2 mm and 0.2 mm and σ = 0.3 mm, 0.3 mm and 0.2 mm, respectively, for the standard mask in the left/right (LR), superior/inferior (SI), and anterior/posterior (AP) translational planes, and Σ = 0.2 mm, 0.1 mm and 0.2 mm and σ = 0.2 mm, 0.3 mm and 0.2 mm, respectively, for the MFD. There was a reduction in rotation errors in the MFD compared to the mask. Margin calculations suggested an isotropic margin could be safely reduced to 2 mm for the MFD.
conclusionThe two devices demonstrate similar daily positional accuracy for fractionated stereotactic treatment of intracranial lesions. Combined with daily image guidance and couch correction, either of these devices is a viable frameless option for fractionated stereotactic radiation therapy.