ArticleJournal of clinical laboratory analysis2016
Analytical Performance Evaluation of Infopia Element™ Auto-coding Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose.
Article in Journal of clinical laboratory analysis, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed.
- Commercial Translation of Electrochemical Biosensors: Supply Chain Strategy, Scale-Up Manufacturing, and Regulatory-Quality Considerations.Biosensors · 2026Review
- Analytical performance of publicly dispensed glucometers in primary health care in a southern Brazilian city.Practical laboratory medicine · 2024Article
- Evaluation of System Accuracy, Precision, Hematocrit Influence, and User Performance of Two Blood Glucose Monitoring Systems Based on ISO 15197:2013/EN ISO 15197:2015.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2024Article
- Assessment of System Accuracy, Intermediate Measurement Precision, and Measurement Repeatability of a Blood Glucose Monitoring System Based on ISO 15197.Journal of diabetes science and technology · 2019Article
- Evaluation of Analytical Performance of Three Blood Glucose Monitoring Systems: System Accuracy, Measurement Repeatability, and Intermediate Measurement Precision.Journal of diabetes science and technology · 2019Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundElement™ Auto-coding Blood Glucose Monitoring System (BGMS; Infopia Co., Ltd., Anyang-si, Korea) was developed for self-monitoring of blood glucose (SMBG).
methodsPrecision, linearity, and interference were tested. Eighty-four capillary blood samples measured by Element™ BGMS were compared with central laboratory method (CLM) results in venous serum. Accuracy was evaluated using ISO 15197:2013 criteria.
resultsCoefficients of variation (CVs; mean) were 2.4% (44.2 mg/dl), 3.7% (100.6 mg/dl), and 2.1% (259.8 mg/dl). Linearity was shown at concentrations 39.25-456.25 mg/l (y = 0.989 + 0.984x, SE = 17.63). Up to 15 mg/dl of galactose, ascorbic acid, and acetaminophen, interference > 10.4% was not observed. Element™ BGMS glucose was higher than CLM levels by 3.2 mg/dl (at 200 mg/dl) to 8.2 mg/dl (at 100 mg/dl). The minimum specification for bias (3.3%) was met at 140 and 200 mg/l glucose. In the Clarke and consensus error grids, 100% of specimens were within zone A and B. For Element™ BGMS values, 92.9% (78/84) to 94.0% (79/84) were within a 15 mg/dl (< 100 mg/dl) or 15% (> 100 mg/dl) of the average CLM value.
conclusionElement™ BGMS was considered an appropriate SMBG for home use; however, the positive bias at low-to-mid glucose levels requires further improvement.