ArticleJournal of the American Society for Mass Spectrometry2016
Experimental Characterization of Secular Frequency Scanning in Ion Trap Mass Spectrometers.
Article in Journal of the American Society for Mass Spectrometry, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed.
- Chromatography-Free Analysis of Mixtures Using a Two-Dimensional Mass Spectrometry (2DMS)-Enabled Quadrupole Time-of-Flight (QToF) Analyzer.Analytical chemistry · 2026Article
- Theoretical fractional formulation of a three-dimensional radio frequency ion trap (Paul-trap) for optimum mass separation.European journal of mass spectrometry (Chichester, England) · 2021Article
- Ion trap operational modes for ion/ion reactions yielding high mass-to-charge product ions.International journal of mass spectrometry · 2020Article
- Increasing the Upper Mass/Charge Limit of a Quadrupole Ion Trap for Ion/Ion Reaction Product Analysis via Waveform Switching.Journal of the American Society for Mass Spectrometry · 2019Article
- Precursor and Neutral Loss Scans in an RF Scanning Linear Quadrupole Ion Trap.Journal of the American Society for Mass Spectrometry · 2018Article
- Implementation of Precursor and Neutral Loss Scans on a Miniature Ion Trap Mass Spectrometer and Performance Comparison to a Benchtop Linear Ion Trap.Journal of the American Society for Mass Spectrometry · 2018Article
- Single Analyzer Precursor Ion Scans in a Linear Quadrupole Ion Trap Using Orthogonal Double Resonance Excitation.Journal of the American Society for Mass Spectrometry · 2017Article
- Successive Resonances for Ion Ejection at Arbitrary Frequencies in an Ion Trap.Journal of the American Society for Mass Spectrometry · 2016Article
- Ion Isolation in a Linear Ion Trap Using Dual Resonance Frequencies.Journal of the American Society for Mass Spectrometry · 2016Article
- Multigenerational Broadband Collision-Induced Dissociation of Precursor Ions in a Linear Quadrupole Ion Trap.Journal of the American Society for Mass Spectrometry · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Secular frequency scanning is implemented and characterized using both a benchtop linear ion trap and a miniature rectilinear ion trap mass spectrometer. Separation of tetraalkylammonium ions and those from a mass calibration mixture and from a pesticide mixture is demonstrated with peak widths approaching unit resolution for optimized conditions using the benchtop ion trap. The effects on the spectra of ion trap operating parameters, including waveform amplitude, scan direction, scan rate, and pressure are explored, and peaks at black holes corresponding to nonlinear (higher-order field) resonance points are investigated. Reverse frequency sweeps (increasing mass) on the Mini 12 are shown to result in significantly higher ion ejection efficiency and superior resolution than forward frequency sweeps that decrement mass. This result is accounted for by the asymmetry in ion energy absorption profiles as a function of AC frequency and the shift in ion secular frequency at higher amplitudes in the trap due to higher order fields. We also found that use of higher AC amplitudes in forward frequency sweeps biases ions toward ejection at points of higher order parametric resonance, despite using only dipolar excitation. Higher AC amplitudes also increase peak width and decrease sensitivity in both forward and reverse frequency sweeps. Higher sensitivity and resolution were obtained at higher trap pressures in the secular frequency scan, in contrast to conventional resonance ejection scans, which showed the opposite trend in resolution on the Mini 12. Mass range is shown to be naturally extended in secular frequency scanning when ejecting ions by sweeping the AC waveform through low frequencies, a method which is similar, but arguably superior, to the more usual method of mass range extension using low q resonance ejection. Graphical Abstract ᅟ.