Observational studyThe British journal of general practice : the journal of the Royal College of General Practitioners2016
Accuracy of monitors used for blood pressure checks in English retail pharmacies: a cross-sectional observational study.
Observational study in The British journal of general practice : the journal of the Royal College of General Practitioners, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- Evaluating the Accuracy of Low-Cost Wearable Sensors for Healthcare Monitoring.Micromachines · 2025Article
- Smartphones and Video Cameras: Future Methods for Blood Pressure Measurement.Frontiers in digital health · 2021Review
- Accuracy of blood-pressure monitors owned by patients with hypertension (ACCU-RATE study): a cross-sectional, observational study in central England.The British journal of general practice : the journal of the Royal College of General Practitioners · 2020Observational
- [Lancet Commission on Hypertension Group position statement on the global improvement of accuracy standards for devices that measure blood pressurePosicionamento do Grupo daRevista panamericana de salud publica = Pan American journal of public health · 2020Article
- Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressure.Journal of hypertension · 2020Article
- Article
- Physician-pharmacist collaborative practice and telehealth may transform hypertension management.Journal of human hypertension · 2019Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
Abstract
backgroundFree blood pressure (BP) checks offered by community pharmacies provide a potentially useful opportunity to diagnose and/or manage hypertension, but the accuracy of the sphygmomanometers in use is currently unknown.
aimTo assess the accuracy of validated automatic BP monitors used for BP checks in a UK retail pharmacy chain. DESIGN AND
settingCross-sectional, observational study in 52 pharmacies from one chain in a range of locations (inner city, suburban, and rural) in central England.
methodMonitor accuracy was compared with a calibrated reference device (Omron PA-350), at 50 mmHg intervals across the range 0-300 mmHg (static pressure test), with a difference from the reference monitor of +/- 3 mmHg at any interval considered a failure. The results were analysed by usage rates and length of time in service.
resultsOf 61 BP monitors tested, eight (13%) monitors failed (that is, were >3 mmHg from reference), all of which underestimated BP. Monitor failure rate from the reference monitor of +/- 3 mmHg at any testing interval varied by length of time in use (2/38, 5% <18 months; 4/14, 29% >18 months, P = 0.038) and to some extent, but non-significantly, by usage rates (4/22, 18% in monitors used more than once daily; 2/33, 6% in those used less frequently, P = 0.204).
conclusionBP monitors within a pharmacy setting fail at similar rates to those in general practice. Annual calibration checks for blood pressure monitors are needed, even for new monitors, as these data indicate declining performance from 18 months onwards.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials