Trial reportMedical physics2016
Can conclusions drawn from phantom-based image noise assessments be generalized to in vivo studies for the nonlinear model-based iterative reconstruction method?
Trial report in Medical physics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 8 citations in OpenAlex.
- Noise power spectrum (NPS) in computed tomography: Enabling local NPS measurement without stationarity and ergodicity assumptions.Medical physics · 2024Article
- Experimental measurement of local noise power spectrum (NPS) in photon counting detector-CT (PCD-CT) using a single data acquisition.Medical physics · 2024Article
- Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details?European radiology · 2018Article
- Impact of bowtie filter and object position on the two-dimensional noise power spectrum of a clinical MDCT system.Medical physics · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
Abstract
purposePhantom-based objective image quality assessment methods are widely used in the medical physics community. For a filtered backprojection (FBP) reconstruction-based linear or quasilinear imaging system, the use of this methodology is well justified. Many key image quality metrics acquired with phantom studies can be directly applied to in vivo human subject studies. Recently, a variety of image quality metrics have been investigated for model-based iterative image reconstruction (MBIR) methods and several novel characteristics have been discovered in phantom studies. However, the following question remains unanswered: can certain results obtained from phantom studies be generalized to in vivo animal studies and human subject studies? The purpose of this paper is to address this question.
methodsOne of the most striking results obtained from phantom studies is a novel power-law relationship between noise variance of MBIR (σ(2)) and tube current-rotation time product (mAs): σ(2) ∝ (mAs)(-0.4) [K. Li et al., "Statistical model based iterative reconstruction (MBIR) in clinical CT systems: Experimental assessment of noise performance," Med. Phys. 41, 041906 (15pp.) (2014)]. To examine whether the same power-law works for in vivo cases, experimental data from two types of in vivo studies were analyzed in this paper. All scans were performed with a 64-slice diagnostic CT scanner (Discovery CT750 HD, GE Healthcare) and reconstructed with both FBP and a MBIR method (Veo, GE Healthcare). An Institutional Animal Care and Use Committee-approved in vivo animal study was performed with an adult swine at six mAs levels (10-290). Additionally, human subject data (a total of 110 subjects) acquired from an IRB-approved clinical trial were analyzed. In this clinical trial, a reduced-mAs scan was performed immediately following the standard mAs scan; the specific mAs used for the two scans varied across human subjects and were determined based on patient size and clinical indications. The measurements of σ(2) were performed at different mAs by drawing regions-of-interest (ROIs) in the liver and the subcutaneous fat. By applying a linear least-squares regression, the β values in the power-law relationship σ(2) ∝ (mAs)(-β) were measured for the in vivo data and compared with the value found in phantom experiments.
resultsFor the in vivo swine study, an exponent of β = 0.43 was found for MBIR, and the coefficient of determination (R(2)) for the corresponding least-squares power-law regression was 0.971. As a reference, the β and R(2) values for FBP were found to be 0.98 and 0.997, respectively, from the same study, which are consistent with the well-known σ(2) ∝ (mAs)(-1.0) relationship for linear CT systems. For the human subject study, the measured β values for the MBIR images were 0.41 ± 0.12 in the liver and 0.37 ± 0.12 in subcutaneous fat. In comparison, the β values for the FBP images were 1.04 ± 0.10 in the liver and 0.97 ± 0.12 in subcutaneous fat. The β values of MBIR and FBP obtained from the in vivo studies were found to be statistically equivalent to the corresponding β values from the phantom study within an equivalency interval of [ - 0.1, 0.1] (p < 0.05); across MBIR and FBP, the difference in β was statistically significant (p < 0.05).
conclusionsDespite the nonlinear nature of the MBIR method, the power-law relationship, σ(2) ∝ (mAs)(-0.4), found from phantom studies can be applied to in vivo animal and human subject studies.