Evidence map›Paper›PMID 26810903›Full record

ArticleInternational orthopaedics2016

Clinical significance of corrosion of cemented femoral stems in metal-on-metal hips: a retrieval study.

Harry S Hothi, Reshid Berber, Andreas C Panagiotopoulos, Robert K Whittaker, Camilla Rhead, John A Skinner, Alister J Hart

Abstract read
PubMed Publisher
In one paragraph

Article in International orthopaedics, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Article
  6. Review
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Harry S HothiInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK. h.hothi@ucl.ac.uk.
Reshid BerberInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.
Andreas C PanagiotopoulosInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.
Robert K WhittakerInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.
Camilla RheadInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.
John A SkinnerInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.
Alister J HartInstitute of Orthopaedics and Musculoskeletal Science, University College London and the Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe clinical significance of corrosion of cemented femoral stems is unclear. The purpose of this retrieval study was to: (1) report on corrosion at the stem-cement interface and (2) correlate these findings with clinical data.

methodsWe analysed cemented stems (n = 36) composed of cobalt-chromium (CoCr) and stainless steel (SS) in a series of revised metal-on-metal hips. We performed detailed inspection of each stem to assess the severity of corrosion at the stem-cement interface using a scale of 1 (low) to 5 (severe). We assessed the severity of corrosion at each stem trunnion and measured wear rates at the head taper and bearing surfaces. We used non-parametric tests to determine the significance of differences between the CoCr and SS stems in relation to: (1) pre-revision whole blood Co and Cr metal ion levels, (2) trunnion corrosion, (3) bearing surface wear and (4) taper material loss.

resultsThe corrosion scores of CoCr stems were significantly greater than SS stems (p < 0.01). Virtually all stem trunnions in both alloy groups had minimal evidence of corrosion. The median pre-revision Co levels of implants with CoCr stems were significantly greater than the SS stems (p < 0.01). There was no significant difference in relation to pre-revision Cr levels (p = 0.521). There was no significant difference between the two stem types in relation to bearing wear (p = 0.926) or taper wear (p = 0.148).

conclusionsSevere corrosion of cemented femoral stems is a common finding at our retrieval centre; surgeons should consider corrosion of CoCr stems as a potential source of metal ions when revising a hip.

Indexed as

AdultAgedAged, 80 and overAlloysArthroplasty, Replacement, HipBone CementsChromiumCobaltCorrosionFemaleFemurHip JointHip ProsthesisHumansMaleMetal-on-Metal Joint ProsthesesAlloysBone CementsChromiumCobaltMetalsCemented stemCorrosionMetal-on-metalRetrieval

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.