Assessment of two-dimensional (2D) and three-dimensional (3D) lower limb measurements in adults: Comparison of micro-dose and low-dose biplanar radiographs.
Article in European radiology, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
5 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Cancer and mortality risks of patients with scoliosis from radiation exposure: a systematic review and meta-analysis.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2020Pooled it
- Subject-Specific Modeling of Femoral Torsion Influences the Prediction of Hip Loading During Gait in Asymptomatic Adults.Frontiers in bioengineering and biotechnology · 2021Article
- Biplanar Low-Dose Radiography Is Accurate for Measuring Combined Anteversion After Total Hip Arthroplasty.HSS journal : the musculoskeletal journal of Hospital for Special Surgery · 2020Article
- 3D hindfoot alignment measurements based on low-dose biplanar radiographs: a clinical feasibility study.Skeletal radiology · 2019Article
- Influence of patient axial malpositioning on the trueness and precision of pelvic parameters obtained from 3D reconstructions based on biplanar radiographs.European radiology · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveTo evaluate reliability of 2D and 3D lower limb measurements in adults using micro-dose compared to low-dose biplanar radiographs(BPR). MATERIALS AND
methodsOne hundred patients (mean 54.9 years) were examined twice using micro-dose and low-dose BPR. Length and mechanical axis of lower limbs were measured on the antero-posterior(ap) micro-dose and low-dose images by two independent readers. Femoral and tibial torsions of 50 patients were measured by two independent readers using reconstructed 3D-models based on the micro-dose and low-dose BPR. Intermethod and interreader agreements were calculated using descriptive statistics, intraclass-correlation-coefficient(ICC), and Bland-Altman analysis.
resultsMean interreader-differences on micro-dose were 0.3 cm(range 0-1.0)/ 0.7°(0-2.9) for limb length/axis and 0.4 cm (0-1.0)/0.8°(0-3.3) on low-dose BPR. Mean intermethod-difference was 0.04 cm ± 0.2/0.04° ± 0.6 for limb length/axis. Interreader-ICC for limb length/axis was 0.999/0.991 on micro-dose and 0.999/0.987 on low-dose BPR. Interreader-ICC for micro-dose was 0.879/0.826 for femoral/ tibial torsion, for low-dose BPR was 0.924/0.909. Mean interreader-differences on micro-dose/low-dose BPR were 3°(0-13°)/2°(0°-12°) for femoral and 4°(0-18°)/3°(0°-10°) for tibial torsion. Mean intermethod-difference was -0.1° ± 5.0/-0.4° ± 2.9 for femoral/tibial torsion. Mean dose-area-product was significantly lower (9.9 times;p < 0.001) for micro-dose BPR.
conclusion2D-and 3D-measurements of lower limbs based on micro-dose BPR are reliable and provide a 10-times lower radiation dose. KEY POINTS: • Lower limb length and mechanical axis can be reliably measured with micro-dose. • Femoral and tibial torsion can be reliably assessed with micro-dose. • Micro-dose allows a huge reduction of radiation exposure.