Evidence map›Paper›PMID 26643076›Full record

ArticleJournal of medical systems2016

Evaluation of Electromagnetic Fields in a Hospital for Safe Use of Electronic Medical Equipment.

Kai Ishida, Tomomi Fujioka, Tetsuo Endo, Ren Hosokawa, Tetsushi Fujisaki, Ryoji Yoshino, Minoru Hirose

Abstract read
PubMed Publisher
In one paragraph

Article in Journal of medical systems, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact, top 87% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 13 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 4 institutions in 1 country.

Kai IshidaDivision of Healthcare Informatics, Faculty of Healthcare, Tokyo Healthcare University, Setagaya 3-11-3, Tokyo, 154-8568, Japan. k-ishida@thcu.ac.jp.ORCID http://orcid.org/0000-0001-8835-6412
Tomomi FujiokaAdvanced Technology Development Section, Building Technology Development Department, Technology Center, Taisei Corporation, Nase-cho 344-1, Yokohama, 245-0051, Japan.
Tetsuo EndoAdvanced Technology Development Section, Building Technology Development Department, Technology Center, Taisei Corporation, Nase-cho 344-1, Yokohama, 245-0051, Japan.
Ren HosokawaDepartment of Medical Safety Engineering, School of Allied Health Science, Kitasato University, Kitasato 1-15-1, Sagamihara, 252-0373, Japan.
Tetsushi FujisakiEnvironmental Survey Office, Shinbashi 3-7-7, Tokyo, 105-0004, Japan.
Ryoji YoshinoEnvironmental Survey Office, Shinbashi 3-7-7, Tokyo, 105-0004, Japan.
Minoru HiroseDepartment of Medical Safety Engineering, School of Allied Health Science, Kitasato University, Kitasato 1-15-1, Sagamihara, 252-0373, Japan.
Geological Survey of Japan · JPKitasato University · JPTaisei (Japan) · JPTokyo Health Care University · JP

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Establishment of electromagnetic compatibility is important in use of electronic medical equipment in hospitals. To evaluate the electromagnetic environment, the electric field intensity induced by electromagnetic radiation in broadcasting spectra coming from outside the hospital was measured in a new hospital building before any patients visited the hospital and 6 months after the opening of the hospital. Various incoming radio waves were detected on the upper floors, with no significant difference in measured levels before and after opening of the hospital. There were no cellphone terminal signals before the hospital opened, but these signals were strongly detected at 6 months thereafter. Cellphone base stations signals were strongly detected on the upper floors, but there were no signals at most locations in the basement and in the center of the building on the lower floors. A maximum electrical intensity of 0.28 V/m from cellphone base stations (2.1 GHz) was detected at the south end of the 2nd floor before the hospital opened. This value is lower than the EMC marginal value for general electronic medical equipment specified in IEC 60601-1-2 (3 V/m). Therefore, electromagnetic interference with electronic medical equipment is unlikely in this situation. However, cellphone terminal signals were frequently detected in non-base station signal areas. This is a concern, and understanding signal strength from cellphone base stations at a hospital is important for promotion of greater safety.

Indexed as

Electromagnetic FieldsElectromagnetic RadiationHospitalsCell PhoneElectronics, MedicalHumansCellphoneElectromagnetic environmentImmunityMeasurementRadio wave