Evidence map›Paper›PMID 26512231›Full record

ArticleIEEE transactions on robotics : a publication of the IEEE Robotics and Automation Society2015

Design of an Optically Controlled MR-Compatible Active Needle.

Seok Chang Ryu, Zhan Fan Quek, Je-Sung Koh, Pierre Renaud, Richard J Black, Behzad Moslehi, Bruce L Daniel, Kyu-Jin Cho, Mark R Cutkosky

Abstract read
In one paragraph

Article in IEEE transactions on robotics : a publication of the IEEE Robotics and Automation Society, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.

0numbers the graph read from it
0cells of the map it votes in
14citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

14 citing papers in PubMed.

  1. Article
  2. FBG-based Shape-Sensing to Enable Lateral Deflection Methods of Autonomous Needle Insertion.Proceedings of the ... IEEE/RSJ International Conference on Intelligent Robots and Systems. IEEE/RSJ International Conference on Intelligent Robots and Systems · 2024
    Article
  3. Review
  4. Review
  5. Article
  6. State of the Art and Future Opportunities in MRI-Guided Robot-Assisted Surgery and Interventions.Proceedings of the IEEE. Institute of Electrical and Electronics Engineers · 2022
    Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Simulation and experimental studies in needle-tissue interactions.Journal of clinical monitoring and computing · 2017
    Article
  13. Article
  14. Methods for Improving the Curvature of Steerable Needles in Biological Tissue.IEEE transactions on bio-medical engineering · 2016
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Seok Chang RyuStanford University, Stanford, CA 94305 USA.
Zhan Fan QuekStanford University, Stanford, CA 94305 USA.
Je-Sung KohBiorobotics Laboratory, School of Mechanical and Aerospace Engineering, Seoul National University, Seoul 151-742,Korea.
Pierre RenaudICube Strasbourg University, CNRS, INSA, 67084 Strasbourg France.
Richard J BlackIntelligent Fiber Optic Systems Corporation, Santa Clara, CA 95054 USA.
Behzad MoslehiIntelligent Fiber Optic Systems Corporation, Santa Clara, CA 95054 USA.
Bruce L DanielDepartment of Radiology, Stanford University, Stanford, CA 94305 USA.
Kyu-Jin ChoBiorobotics Laboratory, School of Mechanical and Aerospace Engineering/IAMD, Seoul National University, Seoul 151-742, Korea.
Mark R CutkoskyStanford University, Stanford, CA 94305 USA.

Funding

MRI Methods for Guiding Focused Ultrasound in the BrainP01CA159992 · NCI · STANFORD UNIVERSITY · PI BUTTS-PAULY, KIM · 2011 to 2015
$8.2M
NCI NIH HHS P01 CA159992
6 · The paper itself

Abstract

An active needle is proposed for the development of magnetic resonance imaging (MRI)-guided percutaneous procedures. The needle uses a low-transition-temperature shape memory alloy (LT SMA) wire actuator to produce bending in the distal section of the needle. Actuation is achieved with internal optical heating using laser light transported via optical fibers and side coupled to the LT SMA. A prototype, with a size equivalent to a standard 16-gauge biopsy needle, exhibits significant bending, with a tip deflection of more than 14° in air and 5° in hard tissue. A single-ended optical sensor with a gold-coated tip is developed to measure the curvature independently of temperature. The experimental results in tissue phantoms show that human tissue causes fast heat dissipation from the wire actuator; however, the active needle can compensate for typical targeting errors during prostate biopsy.

Indexed as

Active needlemechanism designmedical robots and systemsoptical actuation and sensing

Identifiers

PMID26512231
PMCPMC4620588

What OpenQuestion holds

Textmetadata
LicenceTDM
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.