Performance evaluation of the Biograph mCT Flow PET/CT system according to the NEMA NU2-2012 standard.
Article in EJNMMI physics, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 53 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
53 citing papers in PubMed.
- Whole-body CD8Nature communications · 2025Trial
- Attenuation Pre-Correction in Iterative PET Image Reconstruction: What is the Statistical Cost?IEEE transactions on radiation and plasma medical sciences · 2026Article
- Autoradiography and preclinical PET studies with radiolabeled asyn-44 and ACI-12589 for imaging α-synuclein.Journal of Parkinson's disease · 2026Article
- Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts.IEEE transactions on radiation and plasma medical sciences · 2026Article
- Harmonisation of standard uptake values in long-axial field-of-view PET/CT: an exploratory study.EJNMMI physics · 2026Article
- Human subject study of a transmission-aided attenuation correction for precision PET neuroimaging.EJNMMI physics · 2025Article
- PET/CT acquisition and processing protocols in the Netherlands.EJNMMI physics · 2025Article
- Calculation of recovery coefficients for partial volume effect correction in PET/CT imaging using a customized anthropomorphic body phantom.Biomedical engineering online · 2025Article
- Ligand-based design of [Nature communications · 2024Article
- Modular GAN: positron emission tomography image reconstruction using two generative adversarial networks.Frontiers in radiology · 2024Article
- Numerical investigation reveals challenges in measuring the contrast recovery coefficients in PET.Physics in medicine and biology · 2023Article
- Optimization of Q.Clear reconstruction for dynamicEJNMMI physics · 2023Article
- A proper protocol for routineEJNMMI physics · 2023Article
- Direct Patlak Reconstruction of [International journal of molecular sciences · 2023Article
- Toward standardization of tau PET imaging corresponding to various tau PET tracers: a multicenter phantom study.Annals of nuclear medicine · 2023Article
- Optimization of BMI-Based Images for Overweight and Obese Patients - Implications on Image Quality, Quantification, and Radiation Dose in Whole BodyNuclear medicine and molecular imaging · 2023Article
- Variability in PET image quality and quantification measured with a permanently filledEJNMMI physics · 2023Article
- Image quality assessment along the one metre axial field-of-view of the total-body Biograph Vision Quadra PET/CT system forEJNMMI physics · 2022Article
- Quantitative [EJNMMI physics · 2022Article
- Validation ofJournal of nuclear medicine : official publication, Society of Nuclear Medicine · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe purpose of the study is to evaluate the physical performance of a Biograph mCT Flow 64-4R PET/CT system (Siemens Healthcare, Germany) and to compare clinical image quality in step-and-shoot (SS) and continuous table motion (CTM) acquisitions.
methodsThe spatial resolution, sensitivity, count rate curves, and Image Quality (IQ) parameters following the National Electrical Manufactures Association (NEMA) NU2-2012 standard were evaluated. For resolution measurements, an (18)F point source inside a glass capillary tube was used. Sensitivity measurements were based on a 70-cm-long polyethylene tube, filled with 4.5 MBq of FDG. Scatter fraction and count rates were measured using a 70-cm-long polyethylene cylinder with a diameter of 20 cm and a line source (1.04 GBq of FDG) inserted axially into the cylinder 4.5 cm off-centered. A NEMA IQ phantom containing six spheres (10- to 37-mm diameter) was used for the evaluation of the image quality. First, a single-bed scan was acquired (NEMA standard), followed by a two-bed scan (4 min each) of the IQ phantom with the image plane containing the spheres centered in the overlap region of the two bed positions. In addition, a scan of the same region in CTM mode was performed with a table speed of 0.6 mm/s. Furthermore, two patient scans were performed in CTM and SS mode. Image contrasts and patient images were compared between SS and CTM acquisitions.
resultsFull Width Half Maximum (FWHM) of the spatial resolution ranged from 4.3 to 7.8 mm (radial distance 1 to 20 cm). The measured sensitivity was 9.6 kcps/MBq, both at the center of the FOV and 10 cm off-center. The measured noise equivalent count rate (NECR) peak was 185 kcps at 29.0 kBq/ml. The scatter fraction was 33.5 %. Image contrast recovery values (sphere-to-background of 8:1) were between 42 % (10-mm sphere) to 79 % (37-mm sphere). The background variability was between 2.1 and 5.3 % (SS) and between 2.4 and 6.9 % (CTM). No significant difference in image quality was observed between SS and CTM mode.
conclusionsThe spatial resolution, sensitivity, scatter fraction, and count rates were in concordance with the published values for the predecessor system, the Biograph mCT. Contrast recovery values as well as image quality obtained in SS and CTM acquisition modes were similar.