ArticleJournal of diabetes science and technology2015
Performance Evaluation of Three Blood Glucose Monitoring Systems Using ISO 15197: 2013 Accuracy Criteria, Consensus and Surveillance Error Grid Analyses, and Insulin Dosing Error Modeling in a Hospital Setting.
Article in Journal of diabetes science and technology, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 17 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
17 citing papers in PubMed, 27 citations in OpenAlex.
- Performance and Safety of a 15-Day Real-Time Continuous Glucose Monitoring System in Chinese Adults With Diabetes: A Multicenter Trial.Journal of diabetes science and technology · 2026Article
- Accuracy of FreeStyle Libre flash glucose monitoring in patients with type 2 diabetes who migrated from highlands to plains.World journal of diabetes · 2024Article
- Flash glucose monitoring system in gestational diabetes: a study of accuracy and usability.Hormones (Athens, Greece) · 2023Article
- Accuracy of the ContourDiabetology international · 2023Article
- Post-Market Surveillance Assessment of the Clinical Accuracy of a Blood Glucose Monitoring System with an Improved Algorithm for Enhanced Product Performance.Journal of diabetes science and technology · 2023Article
- Mean Absolute Relative Difference of Blood Glucose Monitoring Systems and Relationship to ISO 15197.Journal of diabetes science and technology · 2022Article
- Diabetes Complications and Related Comorbidities Impair the Accuracy of FreeStyle Libre, a Flash Continuous Glucose Monitoring System, in Patients with Type 2 Diabetes.Diabetes, metabolic syndrome and obesity : targets and therapy · 2022Article
- Evaluation of the Freestyle Optium Neo H point-of-care device for measuring blood glucose concentrations in sick calves.Journal of veterinary internal medicine · 2020Article
- Utility of point-of-care vs reference laboratory testing for the evaluation of glucose levels.Diabetic medicine : a journal of the British Diabetic Association · 2019Article
- Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.Journal of diabetes science and technology · 2019Review
- Seven-Year Clinical Surveillance Program Demonstrates Consistent MARD Accuracy Performance of a Blood Glucose Test Strip.Journal of diabetes science and technology · 2018Article
- User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria and Calculation of Insulin Dosing Errors.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2018Article
- Retrospective Analysis of Continuous Glucose Monitoring Data With the Surveillance Error Grid Supports Nonadjunctive Dosing Decisions.Journal of diabetes science and technology · 2017Article
- Commentary Regarding Shapiro, "Nonadjunctive Use of Continuous Glucose Monitors for Insulin Dosing: Is It Safe?"Journal of diabetes science and technology · 2017Article
- A Post-Marketing Surveillance Study to Evaluate Performance of the EXIMO™ Blood Glucose Monitoring System.Journal of clinical and diagnostic research : JCDR · 2017Article
- The Quantitative Relationship Between ISO 15197 Accuracy Criteria and Mean Absolute Relative Difference (MARD) in the Evaluation of Analytical Performance of Self-Monitoring of Blood Glucose (SMBG) Systems.Journal of diabetes science and technology · 2016Article
- Advances in Patient Self-Monitoring of Blood Glucose.Journal of diabetes science and technology · 2015Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundBlood glucose monitoring is an essential component of diabetes management. Inaccurate blood glucose measurements can severely impact patients' health. This study evaluated the performance of 3 blood glucose monitoring systems (BGMS), Contour® Next USB, FreeStyle InsuLinx®, and OneTouch® Verio™ IQ, under routine hospital conditions.
methodsVenous blood samples (N = 236) obtained for routine laboratory procedures were collected at a Spanish hospital, and blood glucose (BG) concentrations were measured with each BGMS and with the available reference (hexokinase) method. Accuracy of the 3 BGMS was compared according to ISO 15197:2013 accuracy limit criteria, by mean absolute relative difference (MARD), consensus error grid (CEG) and surveillance error grid (SEG) analyses, and an insulin dosing error model.
resultsAll BGMS met the accuracy limit criteria defined by ISO 15197:2013. While all measurements of the 3 BGMS were within low-risk zones in both error grid analyses, the Contour Next USB showed significantly smaller MARDs between reference values compared to the other 2 BGMS. Insulin dosing errors were lowest for the Contour Next USB than compared to the other systems.
conclusionsAll BGMS fulfilled ISO 15197:2013 accuracy limit criteria and CEG criterion. However, taking together all analyses, differences in performance of potential clinical relevance may be observed. Results showed that Contour Next USB had lowest MARD values across the tested glucose range, as compared with the 2 other BGMS. CEG and SEG analyses as well as calculation of the hypothetical bolus insulin dosing error suggest a high accuracy of the Contour Next USB.