Evidence map›Paper›PMID 26315073›Full record

ArticleChinese medical journal2015

A Retrospective Study of 1,526 Cases of Transcatheter Occlusion of Patent Ductus Arteriosus.

Mei Jin, Yong-Mei Liang, Xiao-Fang Wang, Bao-Jing Guo, Ke Zheng, Yan Gu, Zhen-Yu Lyu

Open access · goldAbstract read
In one paragraph

Article in Chinese medical journal, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
1.4field-weighted citation impact, top 20% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed, 25 citations in OpenAlex.

  1. Trial
  2. Article
  3. Article
  4. Article
  5. Safety and Efficacy of the Nit-OccludJournal of clinical medicine · 2022
    Article
  6. Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Mei JinDepartment of Pediatric Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, China.
Yong-Mei Liang
Xiao-Fang Wang
Bao-Jing Guo
Ke Zheng
Yan Gu
Zhen-Yu Lyu
Beijing Anzhen Hospital · CNCapital Medical University · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPatent ductus arteriosus (PDA) is one of the most common congenital heart diseases and began to get treated by transcatheter occlusion since 1997 in China. Since then, several devices have been invented for occluding PDA. This study aimed to evaluate the technical feasibility, safety, and efficacy of transcatheter occlusion of PDA with different devices.

methodsOne thousand five hundred and twenty-six patients (537 boys, 989 girls) with PDA from January 1997 to September 2014 underwent descending aortogram and transcatheter occlusion procedure. We retrospectively analyzed data of these patients, including gender, age, weight, size and morphology of PDA, and devices used in transcatheter occlusion, outcomes, and postoperational complications.

resultsMedian age and median weight were 4.0 years (range: 0.3-52.0 years old) and 15.3 kg (range: 4.5-91.0 kg), respectively. Mean ductal diameter, aortic ductal diameter, ductal length, and pulmonary artery pressure were 3.50 ± 2.15 mm, 10.08 ± 2.46 mm, 7.49 ± 3.02 mm, and 30.21 ± 17.28 mmHg, respectively. Morphology of PDA assessed by descending aortogram was of type A in 1428 patients, type B in 6 patients, type C in 79 patients, type D in 4 patients, and type E in 9 patients according to the classification of Krichenko. Of all the 1526 patients, 1497 patients underwent transcatheter PDA closure, among which 1492 were successful. Devices used were Amplatzer duct occluder I (ADO I, 1280, 85.8%), Cook detachable coils (116, 7.8%), ADO II (ADO II, 68, 4.6%), muscular VSD occluder (12, 0.8%), and Amplatzer vascular plug (16, 1.0%).

conclusionsExcellent occlusion rates with low complication rates were achieved with all devices regardless of PDA types. With transcatheter occlusion technique and devices developing, more patients with PDA can be treated with transcatheter closure both safely and efficiently.

Indexed as

AdolescentAdultCardiac CatheterizationChildChild, PreschoolDuctus Arteriosus, PatentFemaleHeart Defects, CongenitalHumansInfantInfant, NewbornMaleMiddle AgedProsthesis DesignRetrospective StudiesSeptal Occluder Device

Identifiers

PMID26315073
PMCPMC4733798
OpenAlexW1542918399

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-SA
Read underepoch 390

Registered trials

None linked