ArticleJournal of diabetes science and technology2015
A Multicenter Performance Evaluation of a Blood Glucose Monitoring System in 21 Leading Hospitals in Spain.
Article in Journal of diabetes science and technology, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Mean Absolute Relative Difference of Blood Glucose Monitoring Systems and Relationship to ISO 15197.Journal of diabetes science and technology · 2022Article
- Seven-Year Clinical Surveillance Program Demonstrates Consistent MARD Accuracy Performance of a Blood Glucose Test Strip.Journal of diabetes science and technology · 2018Article
- A Post-Marketing Surveillance Study to Evaluate Performance of the EXIMO™ Blood Glucose Monitoring System.Journal of clinical and diagnostic research : JCDR · 2017Article
- The Quantitative Relationship Between ISO 15197 Accuracy Criteria and Mean Absolute Relative Difference (MARD) in the Evaluation of Analytical Performance of Self-Monitoring of Blood Glucose (SMBG) Systems.Journal of diabetes science and technology · 2016Article
- Requirements for Successful Adoption of a Glucose Measurement System Into a Hospital POC Program.Journal of diabetes science and technology · 2016Article
- Performance of a Blood Glucose Monitoring System in a Point-of-Care Setting.Journal of diabetes science and technology · 2016Article
- Advances in Patient Self-Monitoring of Blood Glucose.Journal of diabetes science and technology · 2015Article
- Performance Evaluation of Three Blood Glucose Monitoring Systems Using ISO 15197: 2013 Accuracy Criteria, Consensus and Surveillance Error Grid Analyses, and Insulin Dosing Error Modeling in a Hospital Setting.Journal of diabetes science and technology · 2015Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundSelf-monitoring of blood glucose is crucial for the effective self-management of diabetes. The present study evaluated the accuracy of the Contour® XT blood glucose monitoring system (BGMS) compared to the reference method in a large multicenter study under routine lab conditions at each hospital site.
methodsThis study was conducted at 21 leading hospitals in Spain using leftover whole blood samples (n = 2100). Samples were tested with the BGMS using 1 commercial strip lot and the local laboratory hexokinase method. BGMS accuracy was assessed and results were compared to ISO 15197:2013 accuracy limit criteria and by using mean absolute relative difference analysis (MARD), consensus (Parkes) error grid (CEG), and surveillance error grid analyses (SEG).
resultsPooled analysis of 2100 measurements from all sites showed that 99.43% of the BGMS results were within the ranges accepted by the accuracy limit criteria. The overall MARD was 3.85%. MARD was 4.47% for glucose concentrations < 70 mg/dL and 3.81% for concentrations of 70-300 mg/dL. In CEG, most results (99.8%) were within zone A ("no effect on clinical action"); the remaining ones (0.2%) were in zone B ("little to no effect on clinical action"). The SEG analysis showed that most of the results (98.4%) were in the "no risk" zone, with the remaining results in the "slight, lower" risk zone.
conclusionsThis is the largest multicenter study of Contour XT BGMS to date, and shows that this BGMS meets the ISO 15197:2013 accuracy limit criteria under local routine conditions in 21 leading Spanish hospitals.