ArticleOpen heart
Low rate of and rapid attention to inappropriate ICD shocks with remote device and rhythm monitoring: a qualitative study.
Article in Open heart. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesInappropriate shocks are unpleasant and painful. We hypothesise that remote monitoring and careful attention to known and incident atrial fibrillation (AF) can reduce inappropriate shocks to a very low level in clinical praxis.
methodsAltogether 259 patients with implantable cardioverter defibrillator implanted for secondary (S, n=113) and primary (P, n=146) prevention were followed via remote monitoring. At implant, 42S (37%) and 54P (37%) patients had known AF.
resultsInappropriate shocks, all but five due to AF, occurred in 7S (6.2%) and 11P (7.5%), and there were only inappropriate shocks in 5/7S and in 8/11P. They occurred in four of 42S (9.5%) with and in three of 71S (4.2%) without known AF, and in seven of 54P (13%) with and in four of 92P (4.3%) without known AF. The median time from shock to action was 5 and 1 day, respectively. Actions were medication with amiodarone, β blockers, β blockers+amiodarone or β blockers+digoxin (n=5), β blockers+insertion of an atrial lead (n=1), replacement of a fractured lead (n=2), reprogramming in combination with β blockers, digoxin or amiodarone (n=4), reprogramming (n=2) and none (n=4). After action, four further inappropriate shocks occurred during more than 2 years of follow-up, all due to AF.
conclusionsInappropriate shocks occurred at a low rate and most often because of AF known at implant. Remote monitoring enabled rapid action, after which few inappropriate shocks occurred over more than 2 years. Attention to known and incident AF was the most important action to reduce inappropriate shocks.