Trial reportClinical orthopaedics and related research2015
Do oxidized zirconium femoral heads reduce polyethylene wear in cemented THAs? A blinded randomized clinical trial.
Trial report in Clinical orthopaedics and related research, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 1 synthesis or guideline pooled it, 21 citations in OpenAlex.
- Comparison between ceramic-on-polyethylene versus metal-on-polyethylene prostheses in Total Hip Arthroplasties: a systematic review and meta-analysis.Revista da Associacao Medica Brasileira (1992) · 2022Pooled it
- Oxidised zirconium heads show a lower linear wear rate than 28-mm metal heads against XLPE: a retrospective analysis.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026Article
- Review
- The Oxinium Arthrogram: A Sign of Oxidized Zirconium Implant Failure.Arthroplasty today · 2021Article
- Liner dissociation leading to catastrophic failure of an Oxinium femoral head.Arthroplasty today · 2019Article
- Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up.Journal of orthopaedics · 2018Article
- CORR Insights®: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.Clinical orthopaedics and related research · 2018Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundCharnley low-friction torque total hip arthroplasty (THA) remains the gold standard in THA. The main cause for failure is wear of the socket. Highly crosslinked polyethylene (HXLPE) has been associated with reduced wear rates. Also, oxidized zirconium has shown in vitro reduced wear rates. However, to our knowledge, there are no data comparing oxidized zirconium femoral heads with metal heads against HXLPE or ultrahigh-molecular-weight polyethylene (UHMWPE) when 22.25-mm bearings were used, which was the same size that performed so well in Charnley-type THAs. QUESTIONS/PURPOSES: We hypothesized that after a minimal 4-year followup (1) use of HXLPE would result in lower radiographic wear than UHMWPE when articulating with a stainless steel head or with an oxidized zirconium head; (2) use of oxidized zirconium would result in lower radiographic wear than stainless steel when articulating with UHMWPE and HXLPE; and (3) there would be no difference in terms of Merle d'Aubigné scores between the bearing couple combinations.
methodsOne hundred patients were randomized to receive cemented THA with either oxidized zirconium or a stainless steel femoral head. UHMWPE was used in the first 50 patients, whereas HXLPE was used in the next 50 patients. There were 25 patients in each of the four bearing couple combinations. All other parameters were identical in both groups. Complete followup was available in 86 of these patients. Femoral head penetration was measured using a validated computer-assisted method dedicated to all-polyethylene sockets. Clinical results were compared between the groups using the Merle d'Aubigné score.
resultsIn the UHMWPE series, the median steady-state penetration rate from 1 year onward was 0.03 mm/year (range, 0.003-0.25 mm/year) in the oxidized zirconium group versus 0.11 mm/year (range, 0.03-0.29 mm/year) in the metal group (difference of medians 0.08, p < 0.001). In the HXLPE series, the median steady-state penetration rate from 1 year onward was 0.02 mm/year (range, -0.32 to 0.07 mm/year) in the oxidized zirconium group versus 0.05 mm/year (range, -0.39 to 0.11 mm/year) in the metal group (difference of medians 0.03, p < 0.001). The Merle d'Aubigné scores were no different between the groups with a median of 18 in each of the groups (range, 16-18).
conclusionsThis study demonstrated femoral head penetration was reduced by oxidized zirconium when compared with metal on both UHMWPE and HXLPE. However, apart the metal-UHMWE group, all other groups had a steady-state penetration rate well below the osteolysis threshold with a low difference between groups that might not be clinically important at this point. Longer-term followup is needed to warrant whether wear reduction will generate less occurrence of osteolysis and aseptic loosening. LEVEL OF EVIDENCE: Level II, therapeutic study.