Evidence map›Paper›PMID 26062795›Full record

ArticleInternational journal of computer assisted radiology and surgery2015

Computation of the change in length of a braided device when deployed in realistic vessel models.

Hector Fernandez, Juan M Macho, Jordi Blasco, Luis San Roman, Werner Mailaender, Luis Serra, Ignacio Larrabide

Abstract read
PubMed Publisher
In one paragraph

Article in International journal of computer assisted radiology and surgery, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed.

  1. PREDICT: Precise deployment of Silk Vista Baby in confined territory: A technical note.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2026
    Article
  2. Article
  3. Verification of software-based preoperative simulation of flow diverters in clinical cases.Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences · 2023
    Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Hector FernandezGalgo Medical S.L., Barcelona, Spain. hector.fernandez@galgomedical.com.ORCID http://orcid.org/0000-0002-9675-0505
Juan M MachoHospital Clinic Provincial de Barcelona, Barcelona, Spain.
Jordi BlascoHospital Clinic Provincial de Barcelona, Barcelona, Spain.
Luis San RomanHospital Clinic Provincial de Barcelona, Barcelona, Spain.
Werner MailaenderAcandis GmbH, Pforzheim, Germany.
Luis SerraGalgo Medical S.L., Barcelona, Spain.
Ignacio LarrabidePladema, CONICET, UNICEN, Tandil, Argentina. larrabide@exa.unicen.edu.ar.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeAn important issue in the deployment of braided stents, such as flow diverters, is the change in length, also known as foreshortening, underwent by the device when is released from the catheter into a blood vessel. The position of the distal end is controlled by the interventionist, but knowing a priori the position of the proximal end of the device is not trivial. In this work, we assess and validate a novel computer method to predict the length that a braided stent will adopt inside a silicon model of an anatomically accurate vessel.

methodsThree-dimensional rotational angiography images of aneurysmatic patients were used to generate surface models of the vessels (3D meshes) and then create accurate silicon models from them. A braided stent was deployed into each silicon model to measure its length. The same stents deployed on the silicon models were virtually deployed on the 3D meshes using the method being evaluated.

resultsThe method was applied to five stent placements on three different silicon models. The length adopted by the real braided device in the silicon models varies between 15 and 30% from the stent length specified by the manufacturer. The final length predicted by the method was within the estimated error of the measured real stent length.

conclusionsThe method provides, in a few seconds, the length of a braided stent deployed inside a vessel, showing an accurate estimation of the final length for the cases studied. This technique could provide useful information for planning the intervention and improve endovascular treatment of intracranial aneurysms in the future.

Indexed as

AngiographyStentsHumansIntracranial AneurysmModels, AnatomicBraided stentsFlow diverterIntracranial aneurysmsLength change prediction

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.