Evidence map›Paper›PMID 26025935›Full record

ArticleJournal of the American Heart Association2015

Transvenous implantable cardioverter-defibrillator lead reliability: implications for postmarket surveillance.

Daniel B Kramer, Laura A Hatfield, Deepa McGriff, Christopher R Ellis, Melanie T Gura, Michelle Samuel, Linda Kallinen Retel, Robert G Hauser

Abstract read
In one paragraph

Article in Journal of the American Heart Association, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Daniel B KramerBeth Israel Deaconess Medical Center, Boston, MA (D.B.K., M.S.) Hebrew SeniorLife Institute for Aging Research, Boston, MA (D.B.K.) Harvard Medical School, Boston, MA (D.B.K., L.A.H.).
Laura A HatfieldHarvard Medical School, Boston, MA (D.B.K., L.A.H.).
Deepa McGriffMinneapolis Heart Institute Foundation, Minneapolis, MN (D.M.G., L.K.R., R.G.H.).
Christopher R EllisVanderbilt University Medical Center, Nashville, TN (C.R.E.).
Melanie T GuraNortheast Ohio Cardiovascular Specialists, Akron, OH (M.T.G.).
Michelle SamuelBeth Israel Deaconess Medical Center, Boston, MA (D.B.K., M.S.).
Linda Kallinen RetelMinneapolis Heart Institute Foundation, Minneapolis, MN (D.M.G., L.K.R., R.G.H.).
Robert G HauserMinneapolis Heart Institute Foundation, Minneapolis, MN (D.M.G., L.K.R., R.G.H.).

Funding

Patient-Centered Outcomes of Implantable Defibrillator Therapy in Older PatientsK23AG045963 · NIA · HEBREW REHABILITATION CENTER FOR AGED · PI KRAMER, DANIEL BRUCE · 2013 to 2016
$714k
NIA NIH HHS K23 AG045963NIA NIH HHS K23AG045963
6 · The paper itself

Abstract

backgroundAs implantable cardioverter-defibrillator technology evolves, clinicians and patients need reliable performance data on current transvenous implantable cardioverter-defibrillator systems. In addition, real-world reliability data could inform postmarket surveillance strategies directed by regulators and manufacturers. METHODS AND

resultsWe evaluated Medtronic Sprint Quattro, Boston Scientific Endotak, and St Jude Medical Durata and Riata ST Optim leads implanted by participating center physicians between January 1, 2006 and September 1, 2012. Our analytic sample of 2653 patients (median age 65, male 73%) included 445 St Jude, 1819 Medtronic, and 389 Boston Scientific leads. After a median of 3.2 years, lead failure was 0.28% per year (95% CI, 0.19 to 0.43), with no statistically significant difference among manufacturers. Simulations based on these results suggest that detecting performance differences among generally safe leads would require nearly 10 000 patients or very long follow-up.

conclusionsCurrently marketed implantable cardioverter-defibrillator leads rarely fail, which may be reassuring to clinicians advising patients about risks and benefits of transvenous implantable cardioverter-defibrillator systems. Regulators should consider the sample size implications when designing comparative effectiveness studies and evaluating new technology for preventing sudden cardiac death.

Indexed as

Defibrillators, ImplantableProduct Surveillance, PostmarketingAgedEquipment FailureEquipment Failure AnalysisFemaleHumansMaleMiddle AgedRetrospective StudiesICD leadsimplantable cardioverter‐defibrillatorspostmarket surveillance

Identifiers

PMID26025935
PMCPMC4599526

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.