ArticleEuropean radiology2016
Impact of image denoising on image quality, quantitative parameters and sensitivity of ultra-low-dose volume perfusion CT imaging.
Article in European radiology, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed.
- An application study of low-dose computed tomography perfusion imaging for evaluation of the efficacy of neoadjuvant chemotherapy for advanced gastric adenocarcinoma.Gastric cancer : official journal of the International Gastric Cancer Association and the Japanese Gastric Cancer Association · 2018Trial
- Automated contrast-to-noise ratio analysis in chest CT: validation of an open-source segmentation approach.Insights into imaging · 2026Article
- Deep Learning Denoising Improves CT Perfusion Image Quality in the Setting of Lower Contrast Dosing: A Feasibility Study.AJNR. American journal of neuroradiology · 2024Article
- AntiHalluciNet: A Potential Auditing Tool of the Behavior of Deep Learning Denoising Models in Low-Dose Computed Tomography.Diagnostics (Basel, Switzerland) · 2023Article
- Fully automated image quality evaluation on patient CT: Multi-vendor and multi-reconstruction study.PloS one · 2022Article
- [Redundancy information-induced image reconstruction for low-dose myocardial perfusion computed tomography].Nan fang yi ke da xue xue bao = Journal of Southern Medical University · 2018Article
- Low-Dose Volume-Perfusion CT of the Brain: Effects of Radiation Dose Reduction on Performance of Perfusion CT Algorithms.Clinical neuroradiology · 2017Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesTo examine the impact of denoising on ultra-low-dose volume perfusion CT (ULD-VPCT) imaging in acute stroke.
methodsSimulated ULD-VPCT data sets at 20 % dose rate were generated from perfusion data sets of 20 patients with suspected ischemic stroke acquired at 80 kVp/180 mAs. Four data sets were generated from each ULD-VPCT data set: not-denoised (ND); denoised using spatiotemporal filter (D1); denoised using quanta-stream diffusion technique (D2); combination of both methods (D1 + D2). Signal-to-noise ratio (SNR) was measured in the resulting 100 data sets. Image quality, presence/absence of ischemic lesions, CBV and CBF scores according to a modified ASPECTS score were assessed by two blinded readers.
resultsSNR and qualitative scores were highest for D1 + D2 and lowest for ND (all p ≤ 0.001). In 25 % of the patients, ND maps were not assessable and therefore excluded from further analyses. Compared to original data sets, in D2 and D1 + D2, readers correctly identified all patients with ischemic lesions (sensitivity 1.0, kappa 1.0). Lesion size was most accurately estimated for D1 + D2 with a sensitivity of 1.0 (CBV) and 0.94 (CBF) and an inter-rater agreement of 1.0 and 0.92, respectively.
conclusionAn appropriate combination of denoising techniques applied in ULD-VPCT produces diagnostically sufficient perfusion maps at substantially reduced dose rates as low as 20 % of the normal scan. KEY POINTS: Perfusion-CT is an accurate tool for the detection of brain ischemias. The high associated radiation doses are a major drawback of brain perfusion CT. Decreasing tube current in perfusion CT increases image noise and deteriorates image quality. Combination of different image-denoising techniques produces sufficient image quality from ultra-low-dose perfusion CT.