Performance of an Optimized Paper-Based Test for Rapid Visual Measurement of Alanine Aminotransferase (ALT) in Fingerstick and Venipuncture Samples.
Trial report in PloS one, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Microfluidic point-of-care testing device for multiplexed detection of liver function blood markers.Biomedical optics express · 2024Article
- Point-of-Care Testing for Hepatitis Viruses: A Growing Need.Life (Basel, Switzerland) · 2023Review
- Microfluidic Paper-based Device for Medicinal Diagnosis.Current topics in medicinal chemistry · 2022Article
- Increasing the packing density of assays in paper-based microfluidic devices.Biomicrofluidics · 2021Review
- Isoniazid Preventive Therapy for People With HIV Who Are Heavy Alcohol Drinkers in High TB-/HIV-Burden Countries: A Risk-Benefit Analysis.Journal of acquired immune deficiency syndromes (1999) · 2018Article
- Using design strategies from microfluidic device patents to support idea generation.Microfluidics and nanofluidics · 2018Article
- The Cost-effectiveness of a Point-of-Care Paper Transaminase Test for Monitoring Treatment of HIV/TB Co-Infected Persons.Open forum infectious diseases · 2017Article
- High-Resolution Microfluidic Paper-Based Analytical Devices for Sub-Microliter Sample Analysis.Micromachines · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundA paper-based, multiplexed, microfluidic assay has been developed to visually measure alanine aminotransferase (ALT) in a fingerstick sample, generating rapid, semi-quantitative results. Prior studies indicated a need for improved accuracy; the device was subsequently optimized using an FDA-approved automated platform (Abaxis Piccolo Xpress) as a comparator. Here, we evaluated the performance of the optimized paper test for measurement of ALT in fingerstick blood and serum, as compared to Abaxis and Roche/Hitachi platforms. To evaluate feasibility of remote results interpretation, we also compared reading cell phone camera images of completed tests to reading the device in real time.
methods96 ambulatory patients with varied baseline ALT concentration underwent fingerstick testing using the paper device; cell phone images of completed devices were taken and texted to a blinded off-site reader. Venipuncture serum was obtained from 93/96 participants for routine clinical testing (Roche/Hitachi); subsequently, 88/93 serum samples were captured and applied to paper and Abaxis platforms. Paper test and reference standard results were compared by Bland-Altman analysis.
findingsFor serum, there was excellent agreement between paper test and Abaxis results, with negligible bias (+4.5 U/L). Abaxis results were systematically 8.6% lower than Roche/Hitachi results. ALT values in fingerstick samples tested on paper were systematically lower than values in paired serum tested on paper (bias -23.6 U/L) or Abaxis (bias -18.4 U/L); a correction factor was developed for the paper device to match fingerstick blood to serum. Visual reads of cell phone images closely matched reads made in real time (bias +5.5 U/L).
conclusionsThe paper ALT test is highly accurate for serum testing, matching the reference method against which it was optimized better than the reference methods matched each other. A systematic difference exists between ALT values in fingerstick and paired serum samples, and can be addressed by application of a correction factor to fingerstick values. Remote reading of this device is feasible.