Evidence map›Paper›PMID 25900968›Full record

ArticleUltrasonic imaging2016

Dual Modality Noncontact Photoacoustic and Spectral Domain OCT Imaging.

Elisabeth Leiss-Holzinger, Johannes Bauer-Marschallinger, Armin Hochreiner, Philipp Hollinger, Thomas Berer

Open access · hybridAbstract read
In one paragraph

Article in Ultrasonic imaging, 2016. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
1.3field-weighted citation impact, top 21% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 12 citations in OpenAlex.

  1. Article
  2. A review of low-cost and portable optical coherence tomography.Progress in biomedical engineering (Bristol, England) · 2021
    Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 1 institution in 1 country.

Elisabeth Leiss-HolzingerResearch Center for Non-Destructive Testing GmbH (RECENDT), Linz, Austria.
Johannes Bauer-MarschallingerResearch Center for Non-Destructive Testing GmbH (RECENDT), Linz, Austria.
Armin HochreinerResearch Center for Non-Destructive Testing GmbH (RECENDT), Linz, Austria.
Philipp HollingerResearch Center for Non-Destructive Testing GmbH (RECENDT), Linz, Austria.
Thomas BererResearch Center for Non-Destructive Testing GmbH (RECENDT), Linz, Austria thomas.berer@recendt.at.
Research Center for Non Destructive Testing (Austria) · AT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

We developed a multimodal imaging system, combining noncontact photoacoustic imaging and optical coherence tomography (OCT). Photoacoustic signals are recorded without contact to the specimens' surface by using an interferometric technique. The interferometer is realized within a fiber-optic network using a fiber laser at 1550 nm as source. The fiber-optic network allows the integration of a fiber-based OCT system operating at a wavelength region around 1310 nm. Light from the fiber laser and the OCT source are multiplexed into one fiber using wavelength-division multiplexing. The same focusing optics is used for both modalities. Back-reflected light from the sample is demultiplexed and guided to the respective imaging systems. As the same optical components are used for OCT and photoacoustic imaging, the obtained images are co-registered intrinsically in lateral direction. Three-dimensional imaging is implemented by hybrid galvanometer and mechanical scanning. To allow fast B-scan measurements, scanning of the interrogation beam along one dimension is executed by a galvanometer scanner. Slow-axis scanning, perpendicular to the fast axis, is performed utilizing a linear translational stage. We demonstrate two-dimensional and three-dimensional imaging on agarose phantoms.

Indexed as

Equipment DesignFiber Optic TechnologyHumansImaging, Three-DimensionalMultimodal ImagingPhantoms, ImagingPhotoacoustic TechniquesSpectrum AnalysisTomography, Optical Coherencegalvanometer scanninginterferometryoptical coherence tomographyphotoacoustic imagingremote sensing

Identifiers

PMID25900968
PMCPMC4702283
OpenAlexW2142065126

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.