System Accuracy Evaluation of Four Systems for Self-Monitoring of Blood Glucose Following ISO 15197 Using a Glucose Oxidase and a Hexokinase-Based Comparison Method.

Manuela Link, Christina Schmid, Stefan Pleus, Annette Baumstark, Delia Rittmeyer, Cornelia Haug, Guido Freckmann

Registry-linked trialAbstract readEvaluation Study
In one paragraph

Article in Journal of diabetes science and technology, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT01789021 (Analytical Performance Evaluation of Blood Glucose Monitoring Systems), which is not on this map. Cited by 17 papers.

0numbers the graph read from it
0cells of the map it votes in
17citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT01789021 nacompletednot on this map

Analytical Performance Evaluation of Blood Glucose Monitoring Systems

TypeinterventionalSponsorInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität UlmRan2013 to 2013Enrolled120ConditionsDiabetesArms4 blood glucose monitoring systems
3 · Its place in the literature

Who cites it

17 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Polysaccharide IV fromFrontiers in endocrinology · 2018
    Article
  9. Introduction of a Novel Smartphone-Coupled Blood Glucose Monitoring System.Journal of diabetes science and technology · 2017
    Article
  10. Article
  11. Article
  12. Article
  13. Comparative Accuracy of 17 Point-of-Care Glucose Meters.Journal of diabetes science and technology · 2017
    Article
  14. Article
  15. Review
  16. Review
  17. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Manuela LinkInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.
Christina SchmidInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.
Stefan PleusInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany stefan.pleus@uni-ulm.de.
Annette BaumstarkInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.
Delia RittmeyerInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.
Cornelia HaugInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.
Guido FreckmannInstitut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH, Ulm, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe standard ISO (International Organization for Standardization) 15197 is widely accepted for the accuracy evaluation of systems for self-monitoring of blood glucose (SMBG). Accuracy evaluation was performed for 4 SMBG systems (Accu-Chek Aviva, ContourXT, GlucoCheck XL, GlucoMen LX PLUS) with 3 test strip lots each. To investigate a possible impact of the comparison method on system accuracy data, 2 different established methods were used.

methodsThe evaluation was performed in a standardized manner following test procedures described in ISO 15197:2003 (section 7.3). System accuracy was assessed by applying ISO 15197:2003 and in addition ISO 15197:2013 criteria (section 6.3.3). For each system, comparison measurements were performed with a glucose oxidase (YSI 2300 STAT Plus glucose analyzer) and a hexokinase (cobas c111) method.

resultsAll 4 systems fulfilled the accuracy requirements of ISO 15197:2003 with the tested lots. More stringent accuracy criteria of ISO 15197:2013 were fulfilled by 3 systems (Accu-Chek Aviva, ContourXT, GlucoMen LX PLUS) when compared to the manufacturer's comparison method and by 2 systems (Accu-Chek Aviva, ContourXT) when compared to the alternative comparison method. All systems showed lot-to-lot variability to a certain degree; 2 systems (Accu-Chek Aviva, ContourXT), however, showed only minimal differences in relative bias between the 3 evaluated lots.

conclusionsIn this study, all 4 systems complied with the evaluated test strip lots with accuracy criteria of ISO 15197:2003. Applying ISO 15197:2013 accuracy limits, differences in the accuracy of the tested systems were observed, also demonstrating that the applied comparison method/system and the lot-to-lot variability can have a decisive influence on accuracy data obtained for a SMBG system.

Indexed as

Glucose OxidaseHexokinaseAdolescentAdultAgedAged, 80 and overBlood GlucoseBlood Glucose Self-MonitoringDiabetes Mellitus, Type 1Diabetes Mellitus, Type 2HumansMaterials TestingMiddle AgedReference StandardsYoung AdultBlood GlucoseGlucose OxidaseHexokinasecomparison measurement methodglucose oxidase methodhexokinase methodISO 15197:2003ISO 15197:2013self-monitoring of blood glucosesystem accuracy

Identifiers

PMID25872967
PMCPMC4667333

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.