ArticleClinical ophthalmology (Auckland, N.Z.)2015
Opacification of hydrophilic intraocular lenses after Descemet stripping automated endothelial keratoplasty.
Article in Clinical ophthalmology (Auckland, N.Z.), 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
11 citing papers in PubMed, 32 citations in OpenAlex.
- Clinical Characteristics of Patients with Intraocular Lens Calcification after Pars Plana Vitrectomy.Diagnostics (Basel, Switzerland) · 2023Article
- Full-Thickness Compressive Corneal Sutures with Removal of Anterior Chamber Air Bubble in the Management of Acute Corneal Hydrops.Vision (Basel, Switzerland) · 2023Article
- Descemet membrane endothelial keratoplasty: Update on preoperative considerations, surgical techniques, and outcomes.Indian journal of ophthalmology · 2022Review
- A narrative review of intraocular lens opacifications: update 2020.Annals of translational medicine · 2020Review
- Physicochemical Analysis of Sediments Formed on the Surface of Hydrophilic Intraocular Lens after Descemet's Stripping Endothelial Keratoplasty.Materials (Basel, Switzerland) · 2020Article
- Pearly white intraocular lens opacification - "Tertiary cataract".Indian journal of ophthalmology · 2020Article
- Opacification of the C-flex 570C intraocular lens after sulcus fixation.International ophthalmology · 2018Article
- [Opacification of a hydrophilic acrylic intraocular lens after DMEK : A material analysis].Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft · 2017Article
- Hydrophilic intraocular lens opacification after posterior lamellar keratoplasty - a material analysis with special reference to optical quality assessment.BMC ophthalmology · 2017Article
- Hydrophilic intraocular lens opacification after repeated intracameral gas injection for Descemet membrane detachment.Oman journal of ophthalmologyArticle
- Intraocular lens opacification after Descemet's stripping automated endothelial keratoplasty.Taiwan journal of ophthalmologyArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeOpacification of hydrophilic acrylic intraocular lenses (IOLs) is an emerging complication following Descemet stripping automated endothelial keratoplasty (DSAEK). We report six cases and review the current literature.
methodsIn this retrospective, noncomparative, observational case series, patients with IOL opacification after previous DSAEK surgery were identified from corneal clinic records. Case notes were reviewed for demographic details, indication for DSAEK, IOL model, incidence of rebubbling, and postoperative course.
resultsSix patients developed IOL opacification after DSAEK. All patients had Fuchs' endothelial dystrophy and had previously received hydrophilic acrylic IOL models. Central anterior IOL opacification was noted in all six cases. Five cases (83%) had required rebubbling due to dislocated graft tissue, and one had an early postoperative intraocular pressure (IOP) rise. Five cases (83%) were managed conservatively, and one case with a failed graft underwent redo DSAEK and IOL exchange.
conclusionRepeated exposure to intracameral air, raised IOP, and other patient influences may be major etiological factors for IOL opacification after DSAEK. We advise avoiding hydrophilic acrylic IOL models in patients who may require future endothelial keratoplasty.