ArticleInternational orthopaedics2015
Influence of implant design on blood metal ion concentrations in metal-on-metal total hip replacement patients.
Article in International orthopaedics, 2015. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 11 citations in OpenAlex.
- The Influence of the Type of Metal-on-Metal Hip Endoprosthesis on the Clinical, Biochemical, and Oxidative Balance Status-A Comparison of Resurfacing and Metaphyseal Implants.Nanomaterials (Basel, Switzerland) · 2025Article
- What is appropriate surveillance for metal-on-metal hip arthroplasty patients?Acta orthopaedica · 2018Article
- Survivorship and clinical outcome of Birmingham hip resurfacing: a minimum ten years' follow-up.International orthopaedics · 2016Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeMost factors considered to affect blood metal ions following metal-on-metal hip replacement are based on hip resurfacing patients. The study aims were to determine which factors affect blood metal ion concentrations following metal-on-metal total hip replacement (MoM THR).
methodsAll unilateral MoM THR patients at one centre with whole-blood cobalt (Co) and chromium (Cr) concentrations measured up to May 2013 were included. Blood sampling was at a mean of 4.5 years (range 1.1-11.8 years) postoperatively.
resultsOf 496 patients (mean age 59.1 years; 52.8% male), blood metal ions >7 μg/l were observed in 9.7% (n = 48). Large femoral head sizes (≥38 mm) had significantly higher (p < 0.0001) blood metal ions than smaller sizes (28/36 mm). Corail-Pinnacle implants produced significantly lower blood metal ions compared to other implant designs (p < 0.01 Co and Cr). Univariate linear regression demonstrated the only significant predictors of both blood Co and Cr concentrations were femoral head size (R(2) = 8.6% Co and R(2) = 3.3% Cr, both p < 0.0001) and implant design (R(2) = 8.8%, p = 0.005 Co and R(2) = 5.1%, p = 0.003 Cr). When the three THR implant design groups (Corail-Pinnacle, Synergy, Other) were analysed separately, femoral head size no longer significantly affected blood metal ions in any of the three implant design groups.
conclusionsImplant design was the most important factor affecting blood metal ion concentrations. We recommend the regularity of follow-up be tailored to survival rates of various MoM THR designs rather than according to femoral head size.